Your browser doesn't support javascript.
loading
A tether-free, moving actuator total artificial heart.
Min, B G; Kim, H C; Lee, S H; Chang, J K; Choi, J W; Kim, J W; Seo, K P; Rho, J R; Ahn, H; Kim, S W.
Afiliação
  • Min BG; Department of Biomedical Engineering, College of Medicine, Seoul National University, Korea.
ASAIO Trans ; 36(3): M249-51, 1990.
Article em En | MEDLINE | ID: mdl-2252669
An electromechanical moving-actuator TAH was developed and implanted for 5 days in a calf experiment. The efficiency and durability were further improved using pendulous motion of the actuator. This new pendulum pump has a circular motion produced by gear mechanisms connected to the motor. It has a height (from valves to housing bottom) of 7 cm and centrally directed inlets and outlets for good fittability. Active filling (-10 mmHg) inside inner blood sacs was produced by the suction effects of the outer sacs attached to a moving actuator. Two acute animal experiments were performed using the pendulum pump with a Coridale sheep (50 kg weight), and a male calf (70 kg weight). A test was made of 30 minutes of tether-free operation of the implanted pendulum pump inside the small chest of the calf. The pump shows the engineering feasibility of the implantable electrical TAH inside a human chest cavity.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Coração Artificial / Hemodinâmica Limite: Animals Idioma: En Ano de publicação: 1990 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Coração Artificial / Hemodinâmica Limite: Animals Idioma: En Ano de publicação: 1990 Tipo de documento: Article