Calf serum constituent fractions influence polyethylene wear and microbial growth in knee simulator testing.
Brandt, Jan-M; Charron, Kory; Zhao, Lin; MacDonald, Steven J; Medley, John B.
Afiliação
  • Brandt JM; Concordia Joint Replacement Group, Concordia Hip and Knee Institute, Winnipeg, Canada. brandtjm@gmx.net
Proc Inst Mech Eng H ; 226(6): 427-40, 2012 Jun.
Article em En | MEDLINE | ID: mdl-22783759
Calf serum lubricants consisting of various polypeptide constituent fractions are routinely used in knee wear simulators as part of the standardized test protocol. Three calf sera (bovine, new-born and alpha) were diluted as per the recommendation of ISO 14243-3 and used in displacement-controlled knee wear simulators to investigate their effects on polyethylene wear. Biochemical analyses included measuring total polypeptide degradation, electrophoretic profiles and low-molecular weight polypeptide concentrations to elucidate their involvement in the wear process. The effects of the various calf sera constituent fractions on microbial growth were also explored. The polyethylene wear rates and the results from the biochemical analyses for the three calf serum lubricants were all found to be statistically significantly different from each other. The lubricant derived from the alpha-calf serum was closest in constituent fractions to human synovial fluid. It also showed the lowest polyethylene wear rate (14.38 +/- 0.85 mm3/million cycles) and the lowest amount of polypeptide degradation (7.77 +/- 3.87%). Furthermore, the alpha-calf serum lubricant was associated with the least amount of change in the electrophoretic profile, the least change in low-molecular weight polypeptide concentration, and the lowest microbial growth in the presence of sodium azide (a microbial inhibitor conventionally used in implant wear testing). Replacing sodium azide with a broad spectrum antibiotic-antimycotic eradicated the microbial growth. Some speculation was entertained regarding the effect of alpha-calf serum on colloid-mediated boundary lubrication. Based on the results, it was recommended that ISO 14243-3 be modified to include guidelines on calf serum constituent fractions that would favour using alpha-calf serum in order to improve the fidelity of the simulation in knee implant wear testing.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Polietilenos / Teste de Materiais / Soro / Lubrificantes / Prótese do Joelho Tipo de estudo: Guideline Limite: Animals / Humans Idioma: En Ano de publicação: 2012 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Polietilenos / Teste de Materiais / Soro / Lubrificantes / Prótese do Joelho Tipo de estudo: Guideline Limite: Animals / Humans Idioma: En Ano de publicação: 2012 Tipo de documento: Article