High average brightness water window source for short-exposure cryomicroscopy.
Opt Lett
; 37(21): 4425-7, 2012 Nov 01.
Article
em En
| MEDLINE
| ID: mdl-23114317
Laboratory water window cryomicroscopy has recently demonstrated similar image quality as synchrotron-based microscopy but still with much longer exposure times, prohibiting the spread to a wider scientific community. Here we demonstrate high-resolution laboratory water window imaging of cryofrozen cells with 10 s range exposure times. The major improvement is the operation of a λ=2.48 nm, 2 kHz liquid nitrogen jet laser plasma source with high spatial and temporal stability at high average brightness >1.5×10(12) ph/(s×sr×µm(2)×line), i.e., close to that of early synchrotrons. Thus, this source enables not only biological x-ray microscopy in the home laboratory but potentially other applications previously only accessible at synchrotron facilities.