[Comparison of dynamic contour tonometry, Goldmann and pneumotonometer in ocular hypertension patients and their relationship to pachymetry and ocular pulse amplitude]. / Comparación de tonómetro de contorno dinámico, Goldmann y neumotonómetro en pacientes con hipertensión ocular y su relación con la paquimetría y amplitud del pulso ocular.
Arch Soc Esp Oftalmol
; 87(12): 401-6, 2012 Dec.
Article
em Es
| MEDLINE
| ID: mdl-23121701
PURPOSE: To determine the relationship between dynamic contour tonometry (DCT), Goldmann applanation tonometry (GAT) and pneumotonometry (PNT) in ocular hypertension patients (OHT) and their relationship to central corneal thickness (CCT) and ocular pulse amplitude (OPA). METHODS: Sixty patients (101 eyes) with intraocular pressure (IOP) ≥21 mmHg using GAT and normal appearing optic nerve heads and normal visual fields were included. The following tests were performed simultaneously during a single visit: IOP using DCT, GAT and PNT, OPA using DCT and CCT using ultrasound pachymetry. We studied the difference IOP between these 3 tonometers using Wilcoxon non-parametric test and the effect of CCT on IOP and OPA, as well as the relationship between OPA and IOP using Spearman correlation coefficient. RESULTS: The median PNT IOP was 24 mmHg (Inter-quartile range [IQR]: 22-26), median GAT IOP was 22 mmHg (IQR: 22-24), and median DCT IOP was 28.2 mmHg (IQR: 24.1-30.7). PNT and DCT had higher IOP values than GAT (median 2 mmHg and 6.2 mmHg, respectively). Mean CCT was 594.5 µm (SD 30.0). GAT IOP and DCT IOP showed an increase with increased corneal thickness (r:0.209; P=.036 and r:0.195; P=.051, respectively). PNT IOP did not change with CCT (r:0.15; P=.12). The median OPA was 4.8 mmHg (IQR: 3.6-6.1), and significantly increased with GAT IOP (r:0,38; P<.001) and with CCT (r:0.287; P=.004). This association was unclear with IOP PNT and IOP DCT (r:0.067; P=.50 and r:0,17, P=.08, respectively). CONCLUSIONS: DCT and PNT IOP values were higher than GAT IOP measurements in ocular hypertension patients. GAT IOP showed a significant increase with increased corneal thickness. Increased OPA seems to correlate with increased CCT and IOP, particularly if GAT is used.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Tonometria Ocular
/
Hipertensão Ocular
/
Paquimetria Corneana
Tipo de estudo:
Diagnostic_studies
/
Evaluation_studies
Limite:
Adult
/
Aged
/
Aged80
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
Es
Ano de publicação:
2012
Tipo de documento:
Article