Radiation dose and image quality of X-ray volume imaging systems: cone-beam computed tomography, digital subtraction angiography and digital fluoroscopy.
Paul, Jijo; Jacobi, Volkmar; Farhang, Mohammad; Bazrafshan, Babak; Vogl, Thomas J; Mbalisike, Emmanuel C.
Afiliação
  • Paul J; Department of Diagnostic and Interventional Radiology, JW Goethe University Hospital-Frankfurt/Main, Theodor-Stern-Kai 7, 60590 Frankfurt/Main, Germany. jijopaul1980@gmail.com
Eur Radiol ; 23(6): 1582-93, 2013 Jun.
Article em En | MEDLINE | ID: mdl-23250112
ABSTRACT

OBJECTIVE:

Radiation dose and image quality estimation of three X-ray volume imaging (XVI) systems.

METHODS:

A total of 126 patients were examined using three XVI systems (groups 1-3) and their data were retrospectively analysed from 2007 to 2012. Each group consisted of 42 patients and each patient was examined using cone-beam computed tomography (CBCT), digital subtraction angiography (DSA) and digital fluoroscopy (DF). Dose parameters such as dose-area product (DAP), skin entry dose (SED) and image quality parameters such as Hounsfield unit (HU), noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were estimated and compared using appropriate statistical tests.

RESULTS:

Mean DAP and SED were lower in recent XVI than its previous counterparts in CBCT, DSA and DF. HU of all measured locations was non-significant between the groups except the hepatic artery. Noise showed significant difference among groups (P < 0.05). Regarding CNR and SNR, the recent XVI showed a higher and significant difference compared to its previous versions. Qualitatively, CBCT showed significance between versions unlike the DSA and DF which showed non-significance.

CONCLUSION:

A reduction of radiation dose was obtained for the recent-generation XVI system in CBCT, DSA and DF. Image noise was significantly lower; SNR and CNR were higher than in previous versions. The technological advancements and the reduction in the number of frames led to a significant dose reduction and improved image quality with the recent-generation XVI system. KEY POINTS • X-ray volume imaging (XVI) systems are increasingly used for interventional radiological procedures. • More modern XVI systems use lower radiation doses compared with earlier counterparts. • Furthermore more modern XVI systems provide higher image quality. • Technological advances reduce radiation dose and improve image quality.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fluoroscopia / Interpretação de Imagem Radiográfica Assistida por Computador / Angiografia Digital / Tomografia Computadorizada de Feixe Cônico Tipo de estudo: Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fluoroscopia / Interpretação de Imagem Radiográfica Assistida por Computador / Angiografia Digital / Tomografia Computadorizada de Feixe Cônico Tipo de estudo: Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2013 Tipo de documento: Article