Your browser doesn't support javascript.
loading
Evaluation of two, commercial, multi-dye, nucleic acid amplification technology tests, for HBV/HCV/HIV-1/HIV-2 and B19V/HAV, for screening blood and plasma for further manufacture.
Müller, M M; Fraile, M I G; Hourfar, M K; Peris, L B; Sireis, W; Rubin, M G; López, E M; Rodriguez, G T; Seifried, E; Saldanha, J; Schmidt, M.
Afiliação
  • Müller MM; German Red Cross Blood Donor Service, Frankfurt Institute of Transfusion Medicine and Immunohematology, Johann Wolfgang Goethe University, Frankfurt, Germany.
Vox Sang ; 104(1): 19-29, 2013 Jan.
Article em En | MEDLINE | ID: mdl-23252689
ABSTRACT

BACKGROUND:

The cobas TaqScreen MPX Test, version 2.0, a multiplex, multi-dye nucleic acid amplification technology (NAT) test from Roche was evaluated by two European Blood Banks, the German Red Cross Blood Donor Service, Frankfurt, Germany and Centro de Hemoterapia y Hemodonación de Castilla y León, Valladolid, Spain. In addition, the cobas TaqScreen DPX Test was evaluated for the simultaneous detection and quantitation of parvovirus B19 and the detection of hepatitis A virus (HAV). STUDY DESIGN AND

METHODS:

The performances of the two tests were evaluated regarding the analytical sensitivity, the reproducibility of the tests using samples containing low concentrations of each virus and cross-contamination using samples containing high titres of virus.

RESULTS:

The analytical sensitivity of the MPX Test, version 2.0, obtained by the German Red Cross Blood Donor Service was 1·1, 3·9 and 43·3 IU/ml for HBV, HCV and HIV-1, respectively. The comparable analytical sensitivity at Centro de Hemoterapia y Hemodonación de Castilla y León was 3·5, 17·6 and 50·6 IU/ml for HBV, HCV and HIV-1, respectively. The analytical sensitivity of the DPX test determined by the German Red Cross Blood Donor Service was 0·6 and 3·8 IU/ml for HAV and B19.

CONCLUSION:

These multiplex and multi-dye blood screening assays represent a flexible NAT screening system for mini-pools between 6 and 96 samples per pool and fulfil all requirements of the European Pharmacopoeia for HCV and B19V testing of plasma for fractionation. The inclusion of a new multi-dye technology means discriminatory assays are no longer required for either test thus improving workflow, turn-around time and minimize the risk of obtaining a reactive result for which the virus cannot be identified.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Infecções por HIV / Hepatite C / Técnicas de Amplificação de Ácido Nucleico / Segurança do Sangue / Hepatite B Tipo de estudo: Diagnostic_studies / Prognostic_studies / Screening_studies Limite: Humans País como assunto: Europa Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Infecções por HIV / Hepatite C / Técnicas de Amplificação de Ácido Nucleico / Segurança do Sangue / Hepatite B Tipo de estudo: Diagnostic_studies / Prognostic_studies / Screening_studies Limite: Humans País como assunto: Europa Idioma: En Ano de publicação: 2013 Tipo de documento: Article