Biomechanical testing of a new plate system for the distal humerus compared to two well-established implants.
Voigt, Christine; Rank, Christina; Waizner, Klaus; Wendlandt, Robert; Unger, Andreas; Schulz, Arndt P; Jürgens, Christian; Renken, Felix.
Afiliação
  • Voigt C; Department of Trauma and Reconstructive Surgery, Friederikenstift Hospital Hannover, Humboldtstraße 5, 30169, Hannover, Germany. christine.voigt@ddh-gruppe.de
Int Orthop ; 37(4): 667-72, 2013 Apr.
Article em En | MEDLINE | ID: mdl-23359099
PURPOSE: A biomechanical study was performed to test the hypothesis that a new anatomically preformed, thinner, soft-tissue protecting plate system for distal humeral fractures (Tifix®-hybridplate [HP]) would show comparable results in the quasi-static and dynamic testings compared to two conventional implants: The 3.5-mm reconstruction plate (RP) providing primary stability with normal bone mineral density (BMD), and a multidirectional locking plate (Tifix(®)-plate [P]) which can be used with poor bone quality. METHODS: The Tifix(®)-HP was developed by the working group. The biomechanical testing was performed on a C2-fracture-model in 24 synthetic humeri. Three groups, each with eight bone-implant-constructs, were analysed in quasi-static and dynamic tests. RESULTS: The quasi-static measurements showed that under extension loading both locking plates (Tifix(®)-P, Tifix(®)-HP) were significantly stiffer than the reconstruction plate, and that the Tifix(®)-HP had a significantly lower stiffness than the two other implants under flexion loading. In the dynamic tests the Tifix(®)-P allowed significantly less fracture motion compared to the Tifix(®)-HP and the reconstruction plate. In an osteopaenic bone model locking plates failed only under much higher dynamic force than the reconstruction plate. The reconstruction plate and the Tifix(®)-P always failed through screw loosening, whereas the newly developed Tifix(®)-HP showed screw loosening in only one third of cases. CONCLUSION: The hypothesis that the newly designed plate system showed comparable results in the quasi-static and dynamic tests compared to the conventional implants with a significantly lower implant volume and thickness was confirmed.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Placas Ósseas / Fixadores Internos / Fixação Interna de Fraturas / Fraturas do Úmero Limite: Humans Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Placas Ósseas / Fixadores Internos / Fixação Interna de Fraturas / Fraturas do Úmero Limite: Humans Idioma: En Ano de publicação: 2013 Tipo de documento: Article