The influence of head size on corrosion and fretting behaviour at the head-neck interface of artificial hip joints.
Dyrkacz, Richard M R; Brandt, Jan-M; Ojo, Olanrewaju A; Turgeon, Thomas R; Wyss, Urs P.
Afiliação
  • Dyrkacz RM; University of Manitoba, Department of Mechanical and Manufacturing Engineering, E2-327 Engineering and Information Technology Complex, 75A Chancellors Circle, Winnipeg, Manitoba, Canada R3T 2N2.
J Arthroplasty ; 28(6): 1036-40, 2013 Jun.
Article em En | MEDLINE | ID: mdl-23528551
ABSTRACT
The primary goal of this study was to determine if head size affects corrosion and fretting behaviour at the head-neck taper interface of modular hip prostheses. Seventy-four implants were retrieved that featured either a 28 mm or a 36 mm head with a metal-on-polyethylene articulation. The bore of the heads and the neck of the stems were divided into eight regions each and graded by three observers for corrosion and fretting damage separately using modified criteria as reported in the literature. The 36 mm head size featured a significant difference in the corrosion head scores (p=0.022) in comparison to the 28 mm heads. This may be attributed to a greater torque acting along the taper interface due to activities of daily living.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Falha de Prótese / Prótese de Quadril Limite: Aged / Female / Humans / Male Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Falha de Prótese / Prótese de Quadril Limite: Aged / Female / Humans / Male Idioma: En Ano de publicação: 2013 Tipo de documento: Article