A 1.5T MRI-conditional 12-lead electrocardiogram for MRI and intra-MR intervention.
Tse, Zion Tsz Ho; Dumoulin, Charles L; Clifford, Gari D; Schweitzer, Jeff; Qin, Lei; Oster, Julien; Jerosch-Herold, Michael; Kwong, Raymond Y; Michaud, Gregory; Stevenson, William G; Schmidt, Ehud J.
Afiliação
  • Tse ZT; The University of Georgia, College of Engineering, Driftmier Engineering Center, 597 D. W. Brook Drive, Athens, GA 30602, USA.
Magn Reson Med ; 71(3): 1336-47, 2014 Mar.
Article em En | MEDLINE | ID: mdl-23580148
ABSTRACT

PURPOSE:

High-fidelity 12-lead electrocardiogram (ECG) is important for physiological monitoring of patients during MR-guided intervention and cardiac MRI. Issues in obtaining noncorrupted ECGs inside MRI include a superimposed magneto-hydro-dynamic voltage, gradient switching-induced voltages, and radiofrequency heating. These problems increase with magnetic field. The aim of this study is to develop and clinically validate a 1.5T MRI-conditional 12-lead ECG system.

METHODS:

The system was constructed with transmission lines to reduce radiofrequency induction and switching circuits to remove induced voltages. Adaptive filters, trained by 12-lead measurements outside MRI and in two orientations inside MRI, were used to remove the magneto-hydro-dynamic voltage. The system was tested on 10 (one exercising) volunteers and four arrhythmia patients.

RESULTS:

Switching circuits removed most imaging-induced voltages (residual noise <3% of the R-wave). Magneto-hydro-dynamic voltage removal provided intra-MRI ECGs that varied by <3.8% from those outside the MRI, preserving the true S-wave to T-wave segment. In premature ventricular contraction (PVC) patients, clean ECGs separated premature ventricular contraction and sinus rhythm beats. Measured heating was <1.5°C. The system reliably acquired multiphase (steady-state free precession) wall-motion-cine and phase-contrast-cine scans, including subjects in whom 4-lead gating failed. The system required a minimum repetition time of 4 ms to allow robust ECG processing.

CONCLUSION:

High-fidelity intra-MRI 12-lead ECG is possible.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fibrilação Atrial / Cirurgia Assistida por Computador / Eletrocardiografia / Imagem por Ressonância Magnética Intervencionista / Técnicas de Imagem de Sincronização Cardíaca Tipo de estudo: Diagnostic_studies Limite: Aged / Animals / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fibrilação Atrial / Cirurgia Assistida por Computador / Eletrocardiografia / Imagem por Ressonância Magnética Intervencionista / Técnicas de Imagem de Sincronização Cardíaca Tipo de estudo: Diagnostic_studies Limite: Aged / Animals / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2014 Tipo de documento: Article