Your browser doesn't support javascript.
loading
Comparison of in-house developed ELISA with HPLC for measurement of hba1c.
Mohsin, Saima Naz; Ghafoor, Farkhanda; Aasim, Muhammad.
Afiliação
  • Mohsin SN; PMRC Research Centre, Shaikh Zayed Medical Complex, Lahore, Pakistan. saima_razi@hotmail.com
J Immunoassay Immunochem ; 34(3): 305-12, 2013.
Article em En | MEDLINE | ID: mdl-23656250
In-house developed ELISA for HbA1c was evaluated in comparison with Bio-Rad D-10 Dual Program (HPLC). % HbA1c was estimated using said methods in 40 blood samples. Intra and inter assay coefficient of variation (CV) were in acceptable range i.e., <3%. HbA1c results obtained by two methods showed good correlation, (y = 0.8942x + 0.7229 R2 = 0.877). Sensitivity, specificity, positive, and negative predictive value for in-house ELISA classifying patients as controlled and uncontrolled diabetes against HPLC were recorded as 87%, 94.1%, 95.0%, and 93.8%, respectively, while accuracy of the test was 90.0%. In-house ELISA for HbA1c is not only accurate, cost effective and simple but also precise comparable to HPLC in deciding glycemic status of patients.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Hemoglobinas Glicadas / Diabetes Mellitus Limite: Female / Humans / Male Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Hemoglobinas Glicadas / Diabetes Mellitus Limite: Female / Humans / Male Idioma: En Ano de publicação: 2013 Tipo de documento: Article