Your browser doesn't support javascript.
loading
A new method for protection from shower embolism during TEVAR on a shaggy aorta.
Wada, Tomoyuki; Anai, Hirofumi; Shuto, Takashi; Sakaguchi, Takeshi; Hongo, Tetsuo; Shuto, Rieko; Miyamoto, Shinji.
Afiliação
  • Wada T; Department of Cardiovascular Surgery, Oita University Hospital, 1-1-1 Idai-ga-oka, Hazama, Yufu City, Oita, 879-5593, Japan, wada@oita-u.ac.jp.
Gen Thorac Cardiovasc Surg ; 63(2): 105-8, 2015 Feb.
Article em En | MEDLINE | ID: mdl-23794197
ABSTRACT
The case of a patient with a thoracoabdominal aortic aneurysm accompanied by a shaggy aorta, in whom embolism was prevented by a graft used in debranching and placement of an extracorporeal shunt during thoracic endovascular aortic repair, called the "block and trap method", is presented. Two-staged operations were performed using Y graft replacement, debranching bypass, and thoracic endovascular aortic repair during which a temporary shunt line with a blood filter was made involving the femoral artery and vein. The method of trapping emboli in a filter in an external shunt appears effective.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Tromboembolia / Aneurisma da Aorta Torácica / Implante de Prótese Vascular / Procedimentos Endovasculares Limite: Aged80 / Humans / Male Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Tromboembolia / Aneurisma da Aorta Torácica / Implante de Prótese Vascular / Procedimentos Endovasculares Limite: Aged80 / Humans / Male Idioma: En Ano de publicação: 2015 Tipo de documento: Article