Your browser doesn't support javascript.
loading
Evaluation of enzymatic BM Test HbA1c on the JCA-BM6010/C and comparison with Bio-Rad Variant II Turbo, Tosoh HLC 723 G8, and AutoLab immunoturbidimetry assay.
Clin Chem Lab Med ; 51(11): 2201-8, 2013 Nov.
Article em En | MEDLINE | ID: mdl-23898021
ABSTRACT

BACKGROUND:

A novel enzymatic HbA1c assay was introduced for use in an automated chemistry analyzer. With this unique method, HbA1c and plasma glucose can be measured from the same EDTA tube. We evaluated the analytical performance of this enzymatic HbA1c assay in a JCA-BM6010/C analyzer and compared the HbA1c values with the results from other widely used methodological instruments.

METHODS:

The imprecision, linearity, carry-over and concordance rate of the enzymatic HbA1c test (BM Test HbA1c) using the JCA-BM6010/C analyzer were evaluated. Three hundred and seventy-seven specimens with HbA1c concentrations from 16 to 133 mmol/mol were used for a comparison study with two high performance liquid chromatography

methods:

Variant II Turbo and Tosoh HLC 723 G8 and the AutoLab Hemoglobin A1c immunoturbidimetry reagent using a Hitachi 7600-110. Forty specimens were used for the glucose method comparison.

RESULTS:

The HbA1c coefficients of variation of the within-run imprecision for low and high levels were 0.6% and 0.4%, respectively. The linearity of the BM Test HbA1c using the JCA-BM6010/C analyzer was excellent in the range between 31 mmol/mol and 143 mmol/mol. The carry-over rate was 0.2%. The relationships between the BM test and the other three methods were 0.916×Tosoh G8+3.644, r=0.986; 0.887×Bio-Rad Variant II+1.896, r=0.972; and 0.941×AutoLab+4.532, r=0.977. The concordance rates using a cut-off of 48 mmol/mol were 91.5% with Tosoh G8, 82.8% with Bio-Rad Variant II, and 91.0% with AutoLab. The simultaneously assayed plasma glucose with HbA1c was 1.002×Routine plasma glucose+0.625, r=1.000

CONCLUSIONS:

The enzymatic BM Test HbA1c in the JCA-BM6010/C analyzer showed excellent precision and linearity, and a minimal carry-over rate. The simultaneously assayed plasma glucose analysis showed good performance.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Análise Química do Sangue / Glicemia / Hemoglobinas Glicadas / Nefelometria e Turbidimetria Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Análise Química do Sangue / Glicemia / Hemoglobinas Glicadas / Nefelometria e Turbidimetria Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2013 Tipo de documento: Article