Paradoxical thromboembolic stroke during extraction of a recalled St Jude Medical Riata defibrillator lead with conductor externalization.
Larsen, Jacob M; Theuns, Dominic A M J; Thøgersen, Anna M.
Afiliação
  • Larsen JM; Department of Cardiology and Center for Cardiovascular Research, Aalborg University Hospital, Hobrovej 18-22, 9000 Aalborg, Denmark.
Europace ; 16(2): 240, 2014 Feb.
Article em En | MEDLINE | ID: mdl-23933851
ABSTRACT
Recalled St Jude Medical Riata leads have a high rate of insulation failures with potential thrombogenic externalized conductors. We report a paradoxical thromboembolic stroke through a persistent foramen ovale during extraction of a lead with externalization. We suggest mandatory transesophageal echocardiography prior to extraction to screen for lead-related thrombosis and routes of right-to-left shunting to reduce risk of thromboembolic complications.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Cardioversão Elétrica / Desfibriladores Implantáveis / Embolia Paradoxal / Acidente Vascular Cerebral / Remoção de Dispositivo / Forame Oval Patente / Recall de Dispositivo Médico Tipo de estudo: Diagnostic_studies / Etiology_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Humans / Male Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Cardioversão Elétrica / Desfibriladores Implantáveis / Embolia Paradoxal / Acidente Vascular Cerebral / Remoção de Dispositivo / Forame Oval Patente / Recall de Dispositivo Médico Tipo de estudo: Diagnostic_studies / Etiology_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Humans / Male Idioma: En Ano de publicação: 2014 Tipo de documento: Article