ABSTRACT

PURPOSE:

The purpose of this study was to evaluate the utility of k-t parallel imaging for accelerating aortic four-dimensional (4D)-flow MRI. The aim was to systematically investigate the impact of different acceleration factors and number of coil elements on acquisition time, image quality and quantification of hemodynamic parameters.

METHODS:

k-t accelerated 4D-flow MRI (spatial/temporal resolution = 2.1 × 2.5 × 2.5 mm/40.0 ms) was acquired in 10 healthy volunteers with acceleration factors R = 3, 5, and 8 using 12- and 32-channel receiver coils. Results were compared with conventional parallel imaging (GRAPPA [generalized autocalibrating partial parallel acquisition], R = 2). Data analysis included radiological grading of three-dimensional blood flow visualization quality as well as quantification of blood flow, velocities and wall shear stress (WSS).

RESULTS:

k-t GRAPPA significantly reduced scan time by 28%, 54%, and 68%, for R = 3, 5, and 8, respectively, while maintaining image quality as demonstrated by overall similar image quality grading. Significant differences in peak WSS (diff12ch = -5.9%, diff32ch = 18.5%) and mean WSS (diff32ch = 13.9%) were found at the descending aorta for both receiver coils for R = 5 (PWSS < 0.04). Peak velocity differed for R=8 at the aortic root (-7.4%) and descending aorta (-12%) with PpeakVelo < 0.03.

CONCLUSION:

k-t GRAPPA acceleration with a 12- or 32-channel receiver coil and an acceleration of 3 or 5 can compete with a standard GRAPPA R = 2 acceleration.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Aorta / Velocidade do Fluxo Sanguíneo / Algoritmos / Interpretação de Imagem Assistida por Computador / Angiografia por Ressonância Magnética / Imageamento Tridimensional / Técnicas de Imagem de Sincronização Cardíaca Tipo de estudo: Diagnostic_studies Limite: Adult / Female / Humans / Male Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Aorta / Velocidade do Fluxo Sanguíneo / Algoritmos / Interpretação de Imagem Assistida por Computador / Angiografia por Ressonância Magnética / Imageamento Tridimensional / Técnicas de Imagem de Sincronização Cardíaca Tipo de estudo: Diagnostic_studies Limite: Adult / Female / Humans / Male Idioma: En Ano de publicação: 2014 Tipo de documento: Article