Your browser doesn't support javascript.
loading
Novel application of digital microfluidics for the detection of biotinidase deficiency in newborns.
Graham, Carrie; Sista, Ramakrishna S; Kleinert, Jairus; Wu, Ning; Eckhardt, Allen; Bali, Deeksha; Millington, David S; Pamula, Vamsee K.
Afiliação
  • Graham C; Advanced Liquid Logic, Inc. (an Illumina Company), PO Box 14025, Research Triangle Park, NC 27709, USA.
Clin Biochem ; 46(18): 1889-91, 2013 Dec.
Article em En | MEDLINE | ID: mdl-24036022
ABSTRACT

OBJECTIVE:

Newborn screening for biotinidase deficiency can be performed using a fluorometric enzyme assay on dried blood spot specimens. As a pre-requisite to the consolidation of different enzymatic assays onto a single platform, we describe here a novel analytical method for detecting biotinidase deficiency using the same digital microfluidic cartridge that has already been demonstrated to screen for five lysosomal storage diseases (Pompe, Fabry, Gaucher, Hurler and Hunter) in a multiplex format.

METHODS:

A novel assay to quantify biotinidase concentration in dried blood spots (DBS) was developed and optimized on the digital microfluidic platform using proficiency testing samples from the Centers for Disease Control and Prevention. The enzymatic assay uses 4-methylumbelliferyl biotin as the fluorogenic substrate. Biotinidase deficiency assays were performed on normal (n=200) and deficient (n=7) newborn DBS specimens.

RESULTS:

Enzymatic activity analysis of biotinidase deficiency revealed distinct separation between normal and affected DBS specimens using digital microfluidics and these results matched the expected activity.

CONCLUSIONS:

This study has demonstrated performance of biotinidase deficiency assays by measurement of 4-methylumbelliferyl product on a digital microfluidic platform. Due to the inherent ease in multiplexing on such a platform, consolidation of other fluorometric assays onto a single cartridge may be realized.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Triagem Neonatal / Deficiência de Biotinidase / Microfluídica / Ensaios Enzimáticos Tipo de estudo: Diagnostic_studies Limite: Humans / Newborn Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Triagem Neonatal / Deficiência de Biotinidase / Microfluídica / Ensaios Enzimáticos Tipo de estudo: Diagnostic_studies Limite: Humans / Newborn Idioma: En Ano de publicação: 2013 Tipo de documento: Article