Your browser doesn't support javascript.
loading
T1 mapping in the rat myocardium at 7 tesla using a modified CINE inversion recovery sequence.
Smit, Henk; Guridi, Ruben Pellicer; Guenoun, Jamal; Poot, Dirk H J; Doeswijk, Gabriela N; Milanesi, Matteo; Bernsen, Monique R; Krestin, Gabriel P; Klein, Stefan; Kotek, Gyula.
Afiliação
  • Smit H; Departments of Medical Informatics and Radiology, Erasmus MC, Rotterdam, The Netherlands.
J Magn Reson Imaging ; 39(4): 901-10, 2014 Apr.
Article em En | MEDLINE | ID: mdl-24123470
ABSTRACT

PURPOSE:

To evaluate the reproducibility and sensitivity of the modified CINE inversion recovery (mCINE-IR) acquisition on rats for measuring the myocardial T1 at 7 Tesla. MATERIALS AND

METHODS:

The recently published mCINE-IR acquisition on humans was applied on rats for the first time, enabling the possibility of translational studies with an identical sequence. Simulations were used to study signal evolution and heart rate dependency. Gadolinium phantoms, a heart specimen and a healthy rat were used to study reproducibility. Two cryo-infarcted rats were scanned to measure late gadolinium enhancement (LGE).

RESULTS:

In the phantom reproducibility studies the T1 measurements had a maximum coefficient of variation (COV) of 1.3%. For the in vivo reproducibility the COV was below 5% in the anterior cardiac segments. In simulations with phantoms and specimens, a heart rate dependency of approximately 0.5 ms/bpm was present. The T1 maps of the cryo-infarcted rats showed a clear lowering of T1 in de LGE region.

CONCLUSION:

The results show that mCINE-IR is highly reproducible and that the sensitivity allows detecting T1 changes in the rat myocardium.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Algoritmos / Interpretação de Imagem Assistida por Computador / Aumento da Imagem / Imagem Cinética por Ressonância Magnética / Infarto do Miocárdio / Miocárdio Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Algoritmos / Interpretação de Imagem Assistida por Computador / Aumento da Imagem / Imagem Cinética por Ressonância Magnética / Infarto do Miocárdio / Miocárdio Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Ano de publicação: 2014 Tipo de documento: Article