Your browser doesn't support javascript.
loading
An improved sample preparation method for the sensitive detection of peptides by MALDI-MS.
Hioki, Yusaku; Kuyama, Hiroki; Hamana, Chikako; Takeyama, Kohei; Tanaka, Koichi.
Afiliação
  • Hioki Y; Koichi Tanaka Laboratory of Advanced Science and Technology (KTLAST), Shimadzu Corporation, Nishinokyo-Kuwabaracho Nakagyo-ku, Kyoto, 604-8511, Japan.
J Mass Spectrom ; 48(11): 1217-23, 2013 Nov.
Article em En | MEDLINE | ID: mdl-24259210
ABSTRACT
We describe here an optimization study of the sample preparation conditions for sensitive detection of peptides by matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS). Among many factors in the conditions, we varied the percent acetonitrile in the peptide solution, the percent acetonitrile in the matrix solution and the α-cyano-4-hydroxycinnamic acid (CHCA) concentration in the matrix solution. CHCA was chosen because it is the most frequently used matrix for analyzing peptides. The well-established dried-droplet method was employed for sample deposition. The examined range of the concentration of CHCA was from 0.01 to 10 mg/ml, and the MeCN content of the solvent for matrix/analyte was 10% to 50%. The indicator for the detection sensitivity was the S/N ratio of the peaks of peptides used. Highly increased sensitivity (100- to 1000-fold) was observed for the optimal CHCA concentration of 0.1 mg/ml in 20% MeCN/0.1% aq. trifluoroacetic acid (TFA), as compared with the conventional concentration (10 mg/ml) in 50% MeCN/0.1% aq. TFA. For example, the limit of detection of human ACTH 18-39 was 10 amol/well for the optimal condition but 10 fmol/well for the conventional condition. The optimal condition (0.1 mg/ml CHCA in 20% MeCN/0.1% aq. TFA) was verified with five model peptides and provided significant improvement in sensitivity (by two to three orders of magnitude) compared with the conventional conditions. Optimizing the CHCA concentration and solvent composition significantly improved the detection sensitivity in the analysis of peptides by MALDI-MS.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Peptídeos / Espectrometria de Massas por Ionização e Dessorção a Laser Assistida por Matriz Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Peptídeos / Espectrometria de Massas por Ionização e Dessorção a Laser Assistida por Matriz Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 2013 Tipo de documento: Article