Preliminary results: prospective clinical study to assess image-based iterative reconstruction for abdominal computed tomography acquired at 2 radiation dose levels.
Pourjabbar, Sarvenaz; Singh, Sarabjeet; Singh, Anand K; Johnston, Rocio P; Shenoy-Bhangle, Anuradha S; Do, Synho; Padole, Atul; Blake, Michael A; Persson, Anders; Kalra, Mannudeep K.
Afiliação
  • Pourjabbar S; From the *Department of Radiology, Massachusetts General Hospital, Boston, MA; and †Center for Medical Image Science and Visualization, Linköpings universitet/US, Linköping, Sweden.
J Comput Assist Tomogr ; 38(1): 117-22, 2014.
Article em En | MEDLINE | ID: mdl-24424560
ABSTRACT

OBJECTIVE:

The objective of this study was to compare image quality for abdominal computed tomographic (CT) images acquired at 200 and 50 mA s and reconstructed with image-based iterative reconstruction. MATERIALS AND

METHODS:

In this institutional review board-approved prospective study, 22 patients (mean [SD] age, 64.3 [14.4] years; male-female ratio, 1210) gave informed consent for acquisition of additional abdominal CT images on 64-slice multi-detector CT (MDCT) (Siemens Definition Flash). Standard-dose images were acquired at 200 quality reference mA s, whereas low-dose images were acquired at 50 mA s (all series 120 kV; 5-mm section thickness; pitch, 0.91). The low-dose images were reconstructed with a nonlinear 3-dimensional iterative image reconstruction (3D-IIR) (SafeCT; MedicVision, Tirat Carmel, Israel) (4 settings, namely, A1, A2, A3, and A4) and were assessed by 3 abdominal radiologists for lesion detection, image noise, and visibility of small structures. CATPHAN 500 was scanned at the respective doses to obtain noise spectral density and modulation transfer function.

RESULTS:

Subjective image noise was unacceptable at 50-mA s filtered back projection and improved to average in 50-mA s A1 and minimal or no noise in 50-mA s A4. However, the visibility of small structures was similar to standard-dose filtered back projection images on 50-mA s A2. Objective image noise was reduced to 66% for the 50-mA s 3D-IIR images (9.08 [2.3]/26.75 [6.8]). The modulation transfer function curve demonstrated resolution improvement in the low-dose images with the 3D-IIR technique, whereas the noise spectral density curve confirmed noise suppression in the 50-mA s 3D-IIR images.

CONCLUSIONS:

Three-dimensional iterative image reconstruction helps to lower image noise without affecting the visibility of small structures at "moderate" settings. Diagnostically acceptable abdominal CT examinations can be acquired at 75% lower-radiation dose with the help of the image-based iterative reconstruction technique.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Radiográfica Assistida por Computador / Tomografia Computadorizada por Raios X Tipo de estudo: Observational_studies Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Radiográfica Assistida por Computador / Tomografia Computadorizada por Raios X Tipo de estudo: Observational_studies Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2014 Tipo de documento: Article