A novel OCT technique to measure in vivo the corneal adhesiveness for sodium carboxymethylcellulose in humans and its validity in the diagnosis of dry eye.
Napoli, Pietro Emanuele; Coronella, Franco; Satta, Giovanni Maria; Zucca, Ignazio Alberto; Fossarello, Maurizio.
Afiliação
  • Napoli PE; Department of Surgical Sciences, Eye Clinic, University of Cagliari, Cagliari, Italy.
  • Coronella F; Department of Surgical Sciences, Eye Clinic, University of Cagliari, Cagliari, Italy.
  • Satta GM; Department of Surgical Sciences, Eye Clinic, University of Cagliari, Cagliari, Italy.
  • Zucca IA; Department of Surgical Sciences, Eye Clinic, University of Cagliari, Cagliari, Italy.
  • Fossarello M; Department of Surgical Sciences, Eye Clinic, University of Cagliari, Cagliari, Italy.
Invest Ophthalmol Vis Sci ; 55(5): 3179-85, 2014 Apr 24.
Article em En | MEDLINE | ID: mdl-24764061
ABSTRACT

PURPOSE:

The purpose of this work was to gather preliminary data on tear film stability, and the adhesive properties of the corneal surface in dry eye patients and control group subjects, using a new, minimally invasive optical coherence tomography (OCT) imaging method.

METHODS:

We screened 85 human subjects for dry eye and classified them in two groups, as dry eye or normal patients. Sodium carboxymethylcellulose (NaCMC) adhesiveness over the central cornea was measured using Fourier domain anterior segment OCT. The corneal adhesiveness for NaCMC was compared between the two groups, correlated with classical tests, and analyzed for diagnostic validity and repeatability.

RESULTS:

The corneal adhesiveness for NaCMC median and mode values was fair (between 1 and 3 minutes) for dry eye subjects (n = 36) and borderline (between 3 and 5 minutes) for control group subjects (n = 49), and was significantly different between the two groups (P < 0.001). Significant correlations were observed between the corneal adhesiveness measures and dry eye patients' symptoms (P < 0.001), Schirmer I test (P < 0.001), ocular surface staining (P < 0.001), and, particularly, fluorescein break-up time (P < 0.001). The area under the receiver operating characteristic curve was 0.94 (P < 0.001), suggesting reliable sensitivity and specificity of OCT imaging. A statistically significant intraclass correlation (ICC) value of 0.99 was found for measurements of corneal adhesiveness on two subsequent days at the same time (P < 0.001).

CONCLUSIONS:

This minimally invasive, novel technique of OCT imaging of the corneal surface following NaCMC drop instillation provides a measure of corneal adhesiveness. This technique may improve the clinician's ability in the understanding and diagnosis of the dry eye syndrome.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Carboximetilcelulose Sódica / Síndromes do Olho Seco / Córnea / Tomografia de Coerência Óptica Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Carboximetilcelulose Sódica / Síndromes do Olho Seco / Córnea / Tomografia de Coerência Óptica Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2014 Tipo de documento: Article