Your browser doesn't support javascript.
loading
[Development and animal tests of a miniaturized electrical chest compression device].
Sheng Wu Yi Xue Gong Cheng Xue Za Zhi ; 31(1): 39-43, 2014 Feb.
Article em Zh | MEDLINE | ID: mdl-24804481
ABSTRACT
This paper introduces the development and animal tests of a miniaturized electrical chest compression device. Based on pulse width modulation technology produced by micro control unit, the device can control the frequency and depth of the compression accurately, as well as perform real-time adjustment. Therefore, it can perform continuous and stable chest compression for long time, which may increase the successful rate of cardiopulmonary resuscitation (CPR). Besides, the device can also produce different types of compression waveforms, including trapezoidal and triangular waveforms. Then, the performance and efficacy of the device was assessed with a rat model of asphyxial cardiac arrest (CA).
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Reanimação Cardiopulmonar Limite: Animals Idioma: Zh Ano de publicação: 2014 Tipo de documento: Article
Buscar no Google