Your browser doesn't support javascript.
loading
Capsule Endoscopy in a Patient with an Implanted CCM System and an Implantable Defibrillator.
Streitner, Florian; Schoene, Nina; Borggrefe, Martin; Kuschyk, Jürgen.
Afiliação
  • Streitner F; 1st Department of Medicine-Cardiology, University Medical Centre Mannheim, 68167 Mannheim, Germany.
  • Schoene N; 1st Department of Medicine-Cardiology, University Medical Centre Mannheim, 68167 Mannheim, Germany.
  • Borggrefe M; 1st Department of Medicine-Cardiology, University Medical Centre Mannheim, 68167 Mannheim, Germany.
  • Kuschyk J; 1st Department of Medicine-Cardiology, University Medical Centre Mannheim, 68167 Mannheim, Germany.
Case Rep Cardiol ; 2011: 275659, 2011.
Article em En | MEDLINE | ID: mdl-24940511
ABSTRACT
Wireless video capsule endoscopy (CE) is a modern diagnostic tool. Because of its use of digital radiofrequency, it is still relatively contraindicated in patients with implanted cardiac devices. We report the case of a patient with an Optimizer III system delivering cardiac contractility modulating signals (CCM) for heart failure therapy and an implantable cardioverter defibrillator (ICD) who underwent CE. No interferences between the devices were found.

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2011 Tipo de documento: Article