Your browser doesn't support javascript.
loading
Update on retinal vessel structure measurement with spectral-domain optical coherence tomography.
Zhu, Tie Pei; Tong, Yu Hua; Zhan, Hai Jing; Ma, Jin.
Afiliação
  • Zhu TP; Eye Hospital of Wenzhou Medical University, No. 618 Fengqi East Road, Jianggan District, Hangzhou, Zhejiang Province, China. Electronic address: zhutiepei@yahoo.com.
  • Tong YH; Quzhou People's Hospital, Quzhou Central Hospital of Zhejiang Chinese Medical University, No. 2 Zhong Lou Di, Quzhou, Zhejiang Province, China. Electronic address: yuhuatong@126.com.
  • Zhan HJ; Quzhou People's Hospital, Quzhou Central Hospital of Zhejiang Chinese Medical University, No. 2 Zhong Lou Di, Quzhou, Zhejiang Province, China.
  • Ma J; State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-Sen University, No.54 Xian Lie South Road, Guangzhou, Guangdong Province, China. Electronic address: zoc_majin@aliyun.com.
Microvasc Res ; 95: 7-14, 2014 Sep.
Article em En | MEDLINE | ID: mdl-24976361
ABSTRACT
This study was conducted to demonstrate a new scan method for retinal vessel structure measurement in a specific region of fundus (zone B) using spectral-domain optical coherence tomography (SD-OCT), and to assess its reliability. One temporal superior retinal vessel pair passing through a concentric ring (zone B), which was defined between half and one disc distance from the optic disc border, was chosen for the measurement using a volume scan in SD-OCT. On the SD-OCT image, retinal arteriolar outer diameter (RAOD), retinal arteriolar lumen diameter (RALD), retinal venular outer diameter (RVOD) and retinal venular lumen diameter (RVLD) were measured. Retinal vessel diameters on color fundus photographs were also analyzed. Fifty-five healthy individuals were recruited to evaluate intraobserver and interobserver reproducibility between the two examiners. The intraobserver intraclass correlation coefficient (ICC) ranged from 0.972 to 0.981, and the interobserver ICC ranged from 0.968 to 0.980. In the Bland-Altman plot, the 95% limits of interobserver agreement for the RAOD, RALD, RVOD and RVLD were -5.60 to 4.84µm, -5.78 to 5.05µm, -7.52 to 5.62µm and -7.10 to 5.63µm, respectively. The retinal arteriolar and venular lumen diameters on the SD-OCT image were close to the mean arteriolar and venular diameters obtained from the color fundus photographs. Volume scan method produced better images of retinal vessels showing the fine structures of the vessel wall, and provided reliable retinal vessel structure measurement in zone B with good repeatability and reproducibility.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Artéria Retiniana / Veia Retiniana / Tomografia de Coerência Óptica Tipo de estudo: Evaluation_studies / Observational_studies / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Artéria Retiniana / Veia Retiniana / Tomografia de Coerência Óptica Tipo de estudo: Evaluation_studies / Observational_studies / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2014 Tipo de documento: Article