Quantification of myocardial blood flow using (201)Tl SPECT and population-based input function.
Koshino, Kazuhiro; Fukushima, Kazuhito; Fukumoto, Masaji; Hori, Yuki; Moriguchi, Tetsuaki; Zeniya, Tsutomu; Nishimura, Yoshihiro; Kiso, Keisuke; Iida, Hidehiro.
Afiliação
  • Koshino K; Department of Investigative Radiology, National Cerebral and Cardiovascular Center Research Institute, 5-7-1 Fujishirodai, Suita, Osaka, 565-8565, Japan, koshino@ncvc.go.jp.
Ann Nucl Med ; 28(9): 917-25, 2014 Nov.
Article em En | MEDLINE | ID: mdl-25049112
ABSTRACT

OBJECTIVES:

Thallium-201 ((201)Tl) single photon emission computed tomography (SPECT) is an important tool in the diagnosis of ischemic heart disease. Absolute quantification of myocardial blood flow (MBF) has the potential to provide more useful information on myocardial perfusion than semi-quantitative assessments. This study aimed to validate the quantification of MBF using (201)Tl cardiac SPECT based on a population-averaged input function (STD-IF) and one-point blood sample technique.

METHODS:

(201)Tl emission and computed tomography (CT)-based attenuation scans were performed on 11 healthy volunteers at rest using a SPECT/CT scanner. Individual input functions (IND-IFs) during the emission scans were based on arterial blood samples. The STD-IF technique was validated as follows (1) optimal time to calibrate a STD-IF was determined to minimize differences between the calibrated STD-IF and the IND-IFs. (2) Tissue time-activity curves (TTACs) were generated based on a single-tissue compartment model for MBFtrue = 0.5, 1.0, 1.5, and 2.0 mL/min/g, a constant distribution volume of 45 mL/mL, and IND-IFs. The pseudo STD-IF for each subject was generated using the leave-one-out technique. Using the optimal calibration time and the pseudo STD-IFs, MBF values were estimated on the TTACs with an autoradiography method. Optimal mid-scan time (MST) with a fixed duration of 20 min was determined to minimize intersubject variation in estimated MBF errors, and (3) Global and regional MBF values estimated with pseudo STD-IFs were compared to those with IND-IFs using the optimal calibration time and MST.

RESULTS:

The optimal calibration time and MST were both 20 min after (201)Tl injection. Global MBF determined using both IND-IFs and pseudo STD-IF showed significant correlations with rate-pressure products, R (2) = 0.645; p < 0.01 and R (2) = 0.303; p < 0.05, respectively. The mean percent error in regional MBF using pseudo STD-IFs was 0.69 ± 7.80 % (-12.80 to 14.25 %). No significant difference was observed between regional MBF values using IND-IFs and pseudo STD-IFs.

CONCLUSION:

This study demonstrated that the proposed technique based on a STD-IF and one-point blood sample provided hemodynamically reasonable global MBF values and the regional MBF values comparable to those with IND-IFs.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Radioisótopos / Tálio / Tomografia Computadorizada de Emissão de Fóton Único / Compostos Radiofarmacêuticos / Circulação Coronária / Coração Tipo de estudo: Prognostic_studies Limite: Adolescent / Adult / Female / Humans / Male Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Radioisótopos / Tálio / Tomografia Computadorizada de Emissão de Fóton Único / Compostos Radiofarmacêuticos / Circulação Coronária / Coração Tipo de estudo: Prognostic_studies Limite: Adolescent / Adult / Female / Humans / Male Idioma: En Ano de publicação: 2014 Tipo de documento: Article