Evaluation of left ventricular ejection fraction using through-time radial GRAPPA.
Aandal, Gunhild; Nadig, Vidya; Yeh, Victoria; Rajiah, Prabhakar; Jenkins, Trevor; Sattar, Abdus; Griswold, Mark; Gulani, Vikas; Gilkeson, Robert C; Seiberlich, Nicole.
Afiliação
  • Aandal G; Radiology, Case Western Reserve University and University Hospitals Case Medical Center, Cleveland, OH, USA. gunhildaandal@yahoo.no.
  • Nadig V; Haraldsplass Deaconess Hospital, Bergen, Norway. gunhildaandal@yahoo.no.
  • Yeh V; Cardiology, MetroHealth Medical Center at Case Western University, Cleveland, OH, USA. lilnadig@yahoo.co.in.
  • Rajiah P; Case Western Reserve University School of Medicine, Cleveland, OH, USA. vxy16@case.edu.
  • Jenkins T; Radiology, Case Western Reserve University and University Hospitals Case Medical Center, Cleveland, OH, USA. Prabhakar.Rajiah@UHhospitals.org.
  • Sattar A; Division of Cardiovascular Medicine, Harrington Heart & Vascular Institute, University Hospitals Case Medical Center, Case Western Reserve University, Cleveland, OH, USA. Trevor.Jenkins@UHhospitals.org.
  • Griswold M; Epidemiology and Biostatistics, Case Western Reserve University, Cleveland, OH, USA. axs879@case.edu.
  • Gulani V; Radiology, Case Western Reserve University and University Hospitals Case Medical Center, Cleveland, OH, USA. mag46@case.edu.
  • Gilkeson RC; Biomedical Engineering, Case Western Reserve University, Room 309 Wickenden Building 2071 Martin Luther King Jr. Drive, Cleveland, OH, 44106-7207, USA. mag46@case.edu.
  • Seiberlich N; Radiology, Case Western Reserve University and University Hospitals Case Medical Center, Cleveland, OH, USA. vxg46@case.edu.
J Cardiovasc Magn Reson ; 16: 79, 2014 Oct 01.
Article em En | MEDLINE | ID: mdl-25315256
ABSTRACT

BACKGROUND:

The determination of left ventricular ejection fraction using cardiovascular magnetic resonance (CMR) requires a steady cardiac rhythm for electrocardiogram (ECG) gating and multiple breathholds to minimize respiratory motion artifacts, which often leads to scan times of several minutes. The need for gating and breathholding can be eliminated by employing real-time CMR methods such as through-time radial GRAPPA. The aim of this study is to compare left ventricular cardiac functional parameters obtained using current gold-standard breathhold ECG-gated functional scans with non-gated free-breathing real-time imaging using radial GRAPPA, and to determine whether scan time or the occurrence of artifacts are reduced when using this real-time approach.

METHODS:

63 patients were scanned on a 1.5T CMR scanner using both the standard cardiac functional examination with gating and breathholding and the real-time method. Total scan durations were noted. Through-time radial GRAPPA was employed to reconstruct images from the highly accelerated real-time data. The blood volume in the left ventricle was assessed to determine the end systolic volume (ESV), end diastolic volume (EDV), and ejection fraction (EF) for both methods, and images were rated for the presence of artifacts and quality of specific image features by two cardiac readers. Linear regression analysis, Bland-Altman plots and two-sided t-tests were performed to compare the quantitative parameters. A two-sample t-test was performed to compare the scan durations, and a two-sample test of proportion was used to analyze the presence of artifacts. For the reviewers´ ratings the Wilcoxon test for the equality of the scores' distributions was employed.

RESULTS:

The differences in EF, EDV, and ESV between the gold-standard and real-time methods were not statistically significant (p-values of 0.77, 0.82, and 0.97, respectively). Additionally, the scan time was significantly shorter for the real-time data collection (p<0.001) and fewer artifacts were reported in the real-time images (p<0.01). In the qualitative image analysis, reviewers marginally preferred the standard images although some features including cardiac motion were equivalently rated.

CONCLUSION:

Real-time functional CMR with through-time radial GRAPPA performed without ECG-gating under free-breathing can be considered as an alternative to gold-standard breathhold cine imaging for the evaluation of ejection fraction in patients.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Volume Sistólico / Imageamento por Ressonância Magnética / Interpretação de Imagem Assistida por Computador / Função Ventricular Esquerda / Técnicas de Imagem de Sincronização Cardíaca Tipo de estudo: Observational_studies / Prognostic_studies / Qualitative_research / Risk_factors_studies Limite: Humans País como assunto: America do norte Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Volume Sistólico / Imageamento por Ressonância Magnética / Interpretação de Imagem Assistida por Computador / Função Ventricular Esquerda / Técnicas de Imagem de Sincronização Cardíaca Tipo de estudo: Observational_studies / Prognostic_studies / Qualitative_research / Risk_factors_studies Limite: Humans País como assunto: America do norte Idioma: En Ano de publicação: 2014 Tipo de documento: Article