Stability of the Helical TomoTherapy Hi·Art II detector for treatment beam irradiations.
Schombourg, Karin; Bochud, François; Moeckli, Raphaël.
Afiliação
  • Schombourg K; Lausanne University Hospital Institute of Radiation Physics. 123@123.com.
J Appl Clin Med Phys ; 15(6): 4897, 2014 Nov 08.
Article em En | MEDLINE | ID: mdl-25493514
ABSTRACT
The Hi·Art II Helical TomoTherapy (HT) unit is equipped with a built-in onboard MVCT detector used for patient imaging and beam monitoring. Our aim was to study the detector stability for treatment beam measurements. We studied the MVCT detector response with the 6 MV photon beam over time, throughout short-term (during an irradiation) and long-term (two times 50 days) periods. Our results show a coefficient of variation ≤ 1% for detector chambers inside the beam (excluding beam gradients) for short- and long-term response of the MVCT detector. Larger variations were observed in beam gradients and an influence of the X-ray target where degradation was found. The results assume that an 'air scan' procedure is performed daily to recalibrate the detector with the imaging beam. On short term, the detector response stability is comparable to other devices. Long-term measure- ments during two 50-day periods show a good reproducibility. 