Comparison of electrical current and ultrasonic device for incision of the septum of the pharyngoesophageal diverticulum in a pig model.
Hondo, Fábio Yuji; Giordano-Nappi, José Humberto; Pessorrusso, Fernanda Cristina Simões; Franco, Matheus Cavalcante; Neves, Carla Zanelatto; Cecconello, Ivan; Sakai, Paulo; Maluf-Filho, Fauze.
Afiliação
  • Hondo FY; Gastrointestinal Endoscopy Unit, Department of Gastroenterology, Medical School of University of São Paulo, São Paulo, Brazil.
  • Giordano-Nappi JH; Gastrointestinal Endoscopy Unit, Department of Gastroenterology, Medical School of University of São Paulo, São Paulo, Brazil.
  • Pessorrusso FC; Gastrointestinal Endoscopy Unit, Department of Gastroenterology, Medical School of University of São Paulo, São Paulo, Brazil.
  • Franco MC; Gastrointestinal Endoscopy Unit, Department of Gastroenterology, Medical School of University of São Paulo, São Paulo, Brazil. mcavalcantefranco@gmail.com.
  • Neves CZ; Gastrointestinal Endoscopy Unit, Department of Gastroenterology, Medical School of University of São Paulo, São Paulo, Brazil.
  • Cecconello I; Gastrointestinal Surgery Division, Department of Gastroenterology, Medical School of University of São Paulo, Av. Dr. Enéas de Carvalho Aguiar, São Paulo, SP, 05403-000, Brazil.
  • Sakai P; Gastrointestinal Endoscopy Unit, Department of Gastroenterology, Medical School of University of São Paulo, São Paulo, Brazil.
  • Maluf-Filho F; Gastrointestinal Surgery Division, Department of Gastroenterology, Medical School of University of São Paulo, Av. Dr. Enéas de Carvalho Aguiar, São Paulo, SP, 05403-000, Brazil. fauze.maluf@terra.com.br.
Surg Endosc ; 29(11): 3409-13, 2015 Nov.
Article em En | MEDLINE | ID: mdl-25669640
ABSTRACT

BACKGROUND:

Complications of endoscopic treatment of pharyngoesophageal diverticulum (PED) such as bleeding and perforation occur in up to 15 % of cases. Our objective was to compare incision of the PED septum using the ultrasonic device (group H) with incision using electrical current (group M) in a pig model.

METHODS:

Twenty domestic pigs were divided into groups H and M. The incision of the PED septum was performed using the ultrasonic device or using electrical current, respectively. The procedural time, complication rate, length of the border of the incised septum, and thermal tissue damage were compared between groups.

RESULTS:

Mean time to divide the septum and total procedure time were significantly shorter in group H. Mean length of the border of the incised septum was significantly higher in group H. The lateral thermal spread caused by electrical current (group M) was more intense. Hemorrhage was exclusively observed in group M.

CONCLUSION:

Experimental endoscopic diverticulotomy using ultrasonic device was faster and caused less tissue damage when compared to electrical current.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Esofagoscopia / Divertículo de Zenker / Eletrocirurgia / Procedimentos Cirúrgicos Ultrassônicos Tipo de estudo: Evaluation_studies Limite: Animals Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Esofagoscopia / Divertículo de Zenker / Eletrocirurgia / Procedimentos Cirúrgicos Ultrassônicos Tipo de estudo: Evaluation_studies Limite: Animals Idioma: En Ano de publicação: 2015 Tipo de documento: Article