ABSTRACT

OBJECTIVES:

Chang's method, the most widely used attenuation correction (AC) in brain single-photon emission computed tomography (SPECT), requires delineation of the outer contour of the head. Manual and automatic threshold-based methods are prone to errors due to variability of tracer uptake in the scalp. The present study proposes a new method for fully automated delineation of the head based on stereotactical normalization. The method was validated for SPECT with I-123-ioflupane.

METHODS:

The new method was compared to threshold-based delineation in 62 unselected patients who had received I-123-ioflupane SPECT at one of 3 centres. The impact on diagnostic power was tested for semi-quantitative analysis and visual reading of the SPECT images (six independent readers).

RESULTS:

The two delineation methods produced highly consistent semi-quantitative results. This was confirmed by receiver operating characteristic analyses in which the putamen specific-to-background ratio achieved highest area under the curve with negligible effect of the delineation

method:

0.935 versus 0.938 for stereotactical normalization and threshold-based delineation, respectively. Visual interpretation of DVR images was also not affected by the delineation method.

CONCLUSIONS:

Delineation of the head contour by stereotactical normalization appears useful for Chang AC in I-123-ioflupane SPECT. It is robust and does not require user interaction. KEY POINTS •Chang attenuation correction in brain SPECT requires delineation of the head contour. •Manual and threshold-based methods are prone to errors. •The study proposes a fully-automated method for delineation based on stereotactical normalization. •The method is shown to work reliably in I-123-ioflupane SPECT. •It might improve the workflow of I-123-ioflupane SPECT in everyday patient care.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Encéfalo / Tomografia Computadorizada de Emissão de Fóton Único / Técnicas Estereotáxicas / Proteínas da Membrana Plasmática de Transporte de Dopamina / Cabeça Tipo de estudo: Diagnostic_studies / Guideline / Prognostic_studies Limite: Female / Humans / Male Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Encéfalo / Tomografia Computadorizada de Emissão de Fóton Único / Técnicas Estereotáxicas / Proteínas da Membrana Plasmática de Transporte de Dopamina / Cabeça Tipo de estudo: Diagnostic_studies / Guideline / Prognostic_studies Limite: Female / Humans / Male Idioma: En Ano de publicação: 2015 Tipo de documento: Article