Your browser doesn't support javascript.
loading
[Measurement and application of CO2 spectroscopic parameters near 2.0 µm].
Guang Pu Xue Yu Guang Pu Fen Xi ; 34(11): 2881-6, 2014 Nov.
Article em Zh | MEDLINE | ID: mdl-25752024
The accuracy of absorption spectral parameters is very important for the trace gas measurement based on absorption spectroscopy techniques, especially for the isotopic abundance measurement of gas molecules. For some of the applications, spectral parameters listed in HITRAN database were used to retrieve the trace gas concentration. However, these parameters have uncertainty, in order to validate spectroscopic parameters near 2.0 µm of CO2 lines, which are to be used in detecting the CO2 concentration and isotopic abundance, spectra of those lines were recorded at room temperature using a distributed feed-back (DFB) diode laser. The recorded absorption spectra were fitted to Voigt profile. Line position, intensity, self-broadening coefficient and N2-broadening coefficient were deduced from those data. The results show a good consistency in comparison with those listed in HITRAN2012 database. The discrepancy of most line intensities and self-broadening coefficients are less than 2%. The CO2 concentration and Δ(13 CO2 ) in real atmosphere inside laboratory are 440 ppm and -9 per hundred respectively. These results provide a reliable basis for real time and on line detecting the CO2 concentration and Δ(13 CO2) in the wavelength range.
Buscar no Google
Base de dados: MEDLINE Idioma: Zh Ano de publicação: 2014 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Idioma: Zh Ano de publicação: 2014 Tipo de documento: Article