Compensation of spectral and RF errors in swept-source OCT for high extinction complex demodulation.
Opt Express
; 23(5): 5508-20, 2015 Mar 09.
Article
em En
| MEDLINE
| ID: mdl-25836784
We provide a framework for compensating errors within passive optical quadrature demodulation circuits used in swept-source optical coherence tomography (OCT). Quadrature demodulation allows for detection of both the real and imaginary components of an interference fringe, and this information separates signals from positive and negative depth spaces. To achieve a high extinction (â¼60 dB) between these positive and negative signals, the demodulation error must be less than 0.1% in amplitude and phase. It is difficult to construct a system that achieves this low error across the wide spectral and RF bandwidths of high-speed swept-source systems. In a prior work, post-processing methods for removing residual spectral errors were described. Here, we identify the importance of a second class of errors originating in the RF domain, and present a comprehensive framework for compensating both spectral and RF errors. Using this framework, extinctions >60 dB are demonstrated. A stability analysis shows that calibration parameters associated with RF errors are accurate for many days, while those associated with spectral errors must be updated prior to each imaging session. Empirical procedures to derive both RF and spectral calibration parameters simultaneously and to update spectral calibration parameters are presented. These algorithms provide the basis for using passive optical quadrature demodulation circuits with high speed and wide-bandwidth swept-source OCT systems.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Ondas de Rádio
/
Algoritmos
/
Tomografia de Coerência Óptica
Idioma:
En
Ano de publicação:
2015
Tipo de documento:
Article