Your browser doesn't support javascript.
loading
Increasing the accuracy of orbital reconstruction with selective laser-melted patient-specific implants combined with intraoperative navigation.
Rana, Majeed; Chui, Christopher H K; Wagner, Maximillian; Zimmerer, Ruediger; Rana, Madiha; Gellrich, Nils-Claudius.
Afiliação
  • Rana M; Senior Consultant, Department of Craniomaxillofacial Surgery, Hannover Medical School, Hannover, Germany. Electronic address: rana.majeed@mh-hannover.de.
  • Chui CH; Clinical Fellow, Department of Craniomaxillofacial Surgery, Hannover Medical School, Hannover, Germany.
  • Wagner M; Resident, Department of Craniomaxillofacial Surgery, Hannover Medical School, Hannover, Germany.
  • Zimmerer R; Resident, Department of Craniomaxillofacial Surgery, Hannover Medical School, Hannover, Germany.
  • Rana M; Associate Professor, Department of Craniomaxillofacial Surgery, Hannover Medical School, Hannover, Germany.
  • Gellrich NC; Professor and Department Head, Department of Craniomaxillofacial Surgery, Hannover Medical School, Hannover, Germany.
J Oral Maxillofac Surg ; 73(6): 1113-8, 2015 Jun.
Article em En | MEDLINE | ID: mdl-25981837
ABSTRACT

PURPOSE:

Advances in technology have allowed increasing degrees of accuracy in the treatment of orbital deformities. The purpose of this study was to compare the accuracy of pre-bent titanium mesh (PBTM) and selective laser-melted patient-specific implants (PSIs) in unilateral orbital reconstruction after traumatic injury. The authors hypothesized that selective laser-melted PSIs would more accurately reconstruct the orbit. MATERIALS AND

METHODS:

A retrospective review of 34 cases of primary reconstruction of unilateral orbital fractures treated using selective laser-melted PSIs (group 1, n = 17) or PBTM (group 2, n = 17) was performed. The primary outcome measurements were orbital volume excess and the anterior, medial, and posterior intraorbital angles. The Mann-Whitney U test was used to assess the difference in orbital volume and angular deviation between the 2 groups. The level of statistical significance was set at .05. All P values were 2-sided.

RESULTS:

The comparison of mean values for the 2 groups showed significant differences for the anterior angle (PBTM mean, 11.3; standard deviation [SD], 1.8; PSI mean, 4.1; SD, 0.7; P = .001), but not the medial (PBTM mean, 11.6; SD, 2.0; PSI mean, 8.2; SD, 1.9; P = .170) and posterior (PBTM mean, 10.8, SD, 2.8; PSI mean, 8.2, SD, 1.4; P = .760) angles between the unaffected and reconstructed orbits. The postoperative difference in volume between the unaffected and reconstructed orbits differed significantly between the 2 study groups (PBTM mean, 0.6; SD, 0.1; PSI mean, 0.4; SD, 0.1; P = .029).

CONCLUSION:

The results of this study suggest that complex orbital fractures can be reconstructed with an even higher degree of accuracy with selective laser-melted PSIs than with PBTM.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fraturas Orbitárias / Próteses e Implantes / Desenho de Prótese / Procedimentos de Cirurgia Plástica / Cirurgia Assistida por Computador / Lasers Tipo de estudo: Observational_studies / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fraturas Orbitárias / Próteses e Implantes / Desenho de Prótese / Procedimentos de Cirurgia Plástica / Cirurgia Assistida por Computador / Lasers Tipo de estudo: Observational_studies / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2015 Tipo de documento: Article