First-in-man evaluation of the novel balloon delivery system STENTYS Xposition S for the self-apposing coronary artery stent: impact on longitudinal geographic miss during stenting.
Lu, Huangling; IJsselmuiden, Alexander J J; Grundeken, Maik J; Nassif, Martina; de Vries, Arie G; Weevers, Auke; Scholte, Martijn; Spaargaren, René; Wykrzykowska, Joanna J; Tijssen, Jan G P; de Winter, Robbert J; Koch, Karel T.
Afiliação
  • Lu H; Department of Cardiology, Academic Medical Center - University of Amsterdam, Amsterdam, The Netherlands.
EuroIntervention ; 11(12): 1341-5, 2016 Mar.
Article em En | MEDLINE | ID: mdl-25983027
ABSTRACT

AIMS:

A novel balloon delivery system (BDS) for the self-apposing STENTYS sirolimus-eluting stent (SES) has been developed for highly precise longitudinal stent positioning and deployment. The aim of this first-in-man study is to report the quantitative coronary analysis (QCA) angiography and optical coherence tomography (OCT) results as well as the 30-day clinical outcomes of the STENTYS Xposition S SES. METHODS AND

RESULTS:

We included 25 patients (mean age 66.1±10.7 years) with stable coronary artery disease (24%) or acute coronary syndrome (including STEMI in 40%). The device was successfully placed at the intended site in all cases (100%), without procedural complications. Longitudinal geographic miss (entire lesion length [on QCA] not completely covered by the stent) was not observed. Pre-procedural MLD on QCA was 1.30±0.74 mm and post-procedural MLD was 2.74±0.44 mm, p<0.001 (acute gain 1.44±0.70 mm). OCT analyses showed a low percentage of malapposed stent struts directly post stent placement (2.4%), which further decreased after post-dilatation (0.6%, p=0.013), while mean stent area increased (from 9.7 mm2 to 10.5 mm2, p<0.001). At 30-day clinical follow-up, one (4%) major adverse cardiac event (MACE) was observed. One acute stent thrombosis (ST) occurred immediately post procedure in a STEMI patient which was related to inadequate medication therapy.

CONCLUSIONS:

This first-in-man study demonstrated that the use of the novel STENTYS Xposition S balloon delivery system is feasible with a high technical success rate without longitudinal geographical miss. Stent strut malapposition rate directly after STENTYS placement was low and improved further after post-dilatation.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doença da Artéria Coronariana / Cateterismo Cardíaco / Angioplastia Coronária com Balão / Síndrome Coronariana Aguda / Stents Farmacológicos / Cateteres Cardíacos / Infarto do Miocárdio com Supradesnível do Segmento ST Tipo de estudo: Clinical_trials / Etiology_studies / Observational_studies Limite: Aged / Female / Humans / Male / Middle aged País como assunto: Europa Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doença da Artéria Coronariana / Cateterismo Cardíaco / Angioplastia Coronária com Balão / Síndrome Coronariana Aguda / Stents Farmacológicos / Cateteres Cardíacos / Infarto do Miocárdio com Supradesnível do Segmento ST Tipo de estudo: Clinical_trials / Etiology_studies / Observational_studies Limite: Aged / Female / Humans / Male / Middle aged País como assunto: Europa Idioma: En Ano de publicação: 2016 Tipo de documento: Article