Breast cancer detection rates using four different types of mammography detectors.
Mackenzie, Alistair; Warren, Lucy M; Wallis, Matthew G; Cooke, Julie; Given-Wilson, Rosalind M; Dance, David R; Chakraborty, Dev P; Halling-Brown, Mark D; Looney, Padraig T; Young, Kenneth C.
Afiliação
  • Mackenzie A; National Coordinating Centre for the Physics in Mammography (NCCPM), Level B, St Luke's Wing, Royal Surrey County Hospital, Guildford, GU2 7XX, UK. alistairmackenzie@nhs.net.
  • Warren LM; Department of Physics, University of Surrey, Guildford, GU2 7XH, UK. alistairmackenzie@nhs.net.
  • Wallis MG; National Coordinating Centre for the Physics in Mammography (NCCPM), Level B, St Luke's Wing, Royal Surrey County Hospital, Guildford, GU2 7XX, UK.
  • Cooke J; Department of Physics, University of Surrey, Guildford, GU2 7XH, UK.
  • Given-Wilson RM; Cambridge Breast Unit, Cambridge University Hospitals NHS Foundation Trust, Cambridge & NIHR Cambridge Biomedical Research Centre, Cambridge, UK.
  • Dance DR; Jarvis Breast Screening and Diagnostic Centre, Guildford, UK.
  • Chakraborty DP; Department of Radiology, St George's Healthcare NHS Trust, London, UK.
  • Halling-Brown MD; National Coordinating Centre for the Physics in Mammography (NCCPM), Level B, St Luke's Wing, Royal Surrey County Hospital, Guildford, GU2 7XX, UK.
  • Looney PT; Department of Physics, University of Surrey, Guildford, GU2 7XH, UK.
  • Young KC; Department of Radiology, University of Pittsburgh, Pittsburgh, PA, USA.
Eur Radiol ; 26(3): 874-83, 2016 Mar.
Article em En | MEDLINE | ID: mdl-26105023
ABSTRACT

OBJECTIVE:

To compare the performance of different types of detectors in breast cancer detection.

METHODS:

A mammography image set containing subtle malignant non-calcification lesions, biopsy-proven benign lesions, simulated malignant calcification clusters and normals was acquired using amorphous-selenium (a-Se) detectors. The images were adapted to simulate four types of detectors at the same radiation dose digital radiography (DR) detectors with a-Se and caesium iodide (CsI) convertors, and computed radiography (CR) detectors with a powder phosphor (PIP) and a needle phosphor (NIP). Seven observers marked suspicious and benign lesions. Analysis was undertaken using jackknife alternative free-response receiver operating characteristics weighted figure of merit (FoM). The cancer detection fraction (CDF) was estimated for a representative image set from screening.

RESULTS:

No significant differences in the FoMs between the DR detectors were measured. For calcification clusters and non-calcification lesions, both CR detectors' FoMs were significantly lower than for DR detectors. The calcification cluster's FoM for CR NIP was significantly better than for CR PIP. The estimated CDFs with CR PIP and CR NIP detectors were up to 15% and 22% lower, respectively, than for DR detectors.

CONCLUSION:

Cancer detection is affected by detector type, and the use of CR in mammography should be reconsidered. KEY POINTS The type of mammography detector can affect the cancer detection rates. CR detectors performed worse than DR detectors in mammography. Needle phosphor CR performed better than powder phosphor CR. Calcification clusters detection is more sensitive to detector type than other cancers.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Neoplasias da Mama / Calcinose / Mamografia Tipo de estudo: Clinical_trials / Diagnostic_studies / Evaluation_studies / Prognostic_studies / Screening_studies Limite: Aged / Female / Humans / Middle aged Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Neoplasias da Mama / Calcinose / Mamografia Tipo de estudo: Clinical_trials / Diagnostic_studies / Evaluation_studies / Prognostic_studies / Screening_studies Limite: Aged / Female / Humans / Middle aged Idioma: En Ano de publicação: 2016 Tipo de documento: Article