Low Dose Computed Tomography for 3D Planning of Total Hip Arthroplasty: Evaluation of Radiation Exposure and Image Quality.
Huppertz, Alexander; Lembcke, Alexander; Sariali, El-Hadi; Durmus, Tahir; Schwenke, Carsten; Hamm, Bernd; Sparmann, Martin; Baur, Alexander D J.
Afiliação
  • Huppertz A; From the *Department of Radiology, Klinikum Ernst von Bergmann, Potsdam; †Department of Radiology, Charité-Universitätsmedizin Berlin, Berlin, Germany; ‡Department of Orthopaedic Surgery, Hôpital Pitié Salpétrière, Paris, France; §Scossis Statistical Consulting; and ∥Orthopaedic Surgery, Proendo, Berlin, Germany.
J Comput Assist Tomogr ; 39(5): 649-56, 2015.
Article em En | MEDLINE | ID: mdl-26125297
ABSTRACT

OBJECTIVE:

The aim of the study was to compare radiation exposure and image quality between dedicated computed tomography (CT) protocols for preoperative total hip arthroplasty (THA) planning.

METHODS:

Three protocols with automated tube current modulation using 64-slice (n = 177) and 128-slice CT scanners without (n = 129) and with automated tube voltage preselection (n = 84) were compared.

RESULTS:

All 390 CTs were of sufficient quality for THA planning. Mean DLP was 235.0 mGy*cm (effective dose 2.8 mSv). Lowest radiation exposure (2.5 mSv) was seen with automated voltage preselection and the algorithm's selection was 100 kV (90.5% of patients) and 120 kV. Lowest image noise was seen in the highest dose group (3.1 mSv, 128-slice CT fixed tube voltage). A significant difference in cortical bone radiodensity was seen between 100 kV and 120 kV (P < 0.0001).

CONCLUSIONS:

Preoperative pelvic CT for THA planning is possible with very low radiation dose and reliable quality. Automated voltage preselection further decreases the effective dose by 18.2%.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doses de Radiação / Interpretação de Imagem Radiográfica Assistida por Computador / Tomografia Computadorizada por Raios X / Exposição à Radiação / Artroplastia de Quadril Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doses de Radiação / Interpretação de Imagem Radiográfica Assistida por Computador / Tomografia Computadorizada por Raios X / Exposição à Radiação / Artroplastia de Quadril Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2015 Tipo de documento: Article