Hyperpolarized (129) Xe imaging of the rat lung using spiral IDEAL.
Doganay, Ozkan; Wade, Trevor; Hegarty, Elaine; McKenzie, Charles; Schulte, Rolf F; Santyr, Giles E.
Afiliação
  • Doganay O; Department of Medical Biophysics, Western University, London, ON, Canada.
  • Wade T; Imaging Research Laboratories, Robarts Research Institute, London, ON, Canada.
  • Hegarty E; Department of Medical Biophysics, Western University, London, ON, Canada.
  • McKenzie C; Imaging Research Laboratories, Robarts Research Institute, London, ON, Canada.
  • Schulte RF; Imaging Research Laboratories, Robarts Research Institute, London, ON, Canada.
  • Santyr GE; Department of Medical Biophysics, Western University, London, ON, Canada.
Magn Reson Med ; 76(2): 566-76, 2016 08.
Article em En | MEDLINE | ID: mdl-26332385
ABSTRACT

PURPOSE:

To implement and optimize a single-shot spiral encoding strategy for rapid 2D IDEAL projection imaging of hyperpolarized (Hp) (129) Xe in the gas phase, and in the pulmonary tissue (PT) and red blood cells (RBCs) compartments of the rat lung, respectively. THEORY AND

METHODS:

A theoretical and experimental point spread function analysis was used to optimize the spiral k-space read-out time in a phantom. Hp (129) Xe IDEAL images from five healthy rats were used to (i) optimize flip angles by a Bloch equation analysis using measured kinetics of gas exchange and (ii) investigate the feasibility of the approach to characterize the exchange of Hp (129) Xe.

RESULTS:

A read-out time equal to approximately 1.8 × T2* was found to provide the best trade-off between spatial resolution and signal-to-noise ratio (SNR). Spiral IDEAL approaches that use the entire dissolved phase magnetization should give an SNR improvement of a factor of approximately three compared with Cartesian approaches with similar spatial resolution. The IDEAL strategy allowed imaging of gas, PT, and RBC compartments with sufficient SNR and temporal resolution to permit regional gas exchange measurements in healthy rats.

CONCLUSION:

Single-shot spiral IDEAL imaging of gas, PT and RBC compartments and gas exchange is feasible in rat lung using Hp (129) Xe. Magn Reson Med 76566-576, 2016. © 2015 Wiley Periodicals, Inc.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Isótopos de Xenônio / Processamento de Sinais Assistido por Computador / Imageamento por Ressonância Magnética / Interpretação de Imagem Assistida por Computador / Troca Gasosa Pulmonar / Pulmão Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Isótopos de Xenônio / Processamento de Sinais Assistido por Computador / Imageamento por Ressonância Magnética / Interpretação de Imagem Assistida por Computador / Troca Gasosa Pulmonar / Pulmão Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Ano de publicação: 2016 Tipo de documento: Article