Flow Test to Predict Early Hypotony and Hypertensive Phase After Ahmed Glaucoma Valve (AGV) Surgical Implantation.
J Glaucoma
; 25(6): 493-6, 2016 06.
Article
em En
| MEDLINE
| ID: mdl-26372153
ABSTRACT
PURPOSE:
To assess the validity of a preimplantation flow test to predict early hypotony [intraocular pressure (IOP)≤5 mm Hg on 2 consecutive visits and hypertensive phase (HP) (IOP>21 mm Hg) after Ahmed Glaucoma Valve (AGV) implantation. PATIENTS ANDMETHODS:
Prospective interventional study on patients receiving an AGV. A preimplantation flow test using a gravity-driven reservoir and an open manometer was performed on all AGVs. Opening pressure (OP) and closing pressure (CP) were defined as the pressure at which fluid was seen to flow or stop flowing through the AGV, respectively. OP and CP were measured twice per AGV. Patients were followed for 12 weeks.RESULTS:
In total, 20 eyes from 19 patients were enrolled. At 12 weeks the mean IOP decreased from 29.2±9.1 to 16.8±5.2 mm Hg (P<0.01). The mean AGV OP was 17.5±5.4 mm Hg and the mean CP was 6.7±2.3 mm Hg. Early (within 2 wk postoperative) HP occurred in 37% and hypotony in 16% of cases. An 18 mm Hg cutoff for the OP gave a sensitivity of 0.71, specificity of 0.83, positive predictive value of 0.71, and negative predictive value of 0.83 for predicting an early HP. A 7 mm Hg cutoff for the CP yielded a sensitivity of 1.0, specificity of 0.38, positive predictive value of 0.23, and negative predictive value of 1.0 for predicting hypotony.CONCLUSIONS:
Preoperative OP and CP may predict early hypotony or HP and may be used as a guide as to which AGV valves to discard before implantation surgery.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Humor Aquoso
/
Hipotensão Ocular
/
Glaucoma
/
Hipertensão Ocular
/
Implantes para Drenagem de Glaucoma
/
Técnicas de Diagnóstico Oftalmológico
/
Pressão Intraocular
Tipo de estudo:
Diagnostic_studies
/
Observational_studies
/
Prognostic_studies
/
Risk_factors_studies
Limite:
Aged
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Ano de publicação:
2016
Tipo de documento:
Article