Your browser doesn't support javascript.
loading
Calibrating oscillation response of a piezo-stage using optical tweezers.
Opt Express ; 23(19): 24108-19, 2015 Sep 21.
Article em En | MEDLINE | ID: mdl-26406617
ABSTRACT
In optical tweezers, a piezo-stage (PZT) is widely used to precisely position samples for force clamp, calibrating optical trap and stretching DNA. For a trapped bead in solution, the oscillation response of PZT is vital for all kinds of applications. A coupling ratio, actual amplitude to nominal amplitude, can be calibrated by power spectral density during sinusoidal oscillations. With oscillation frequency increasing, coupling ratio decreases in both x- and y-directions, which is also confirmed by the calibration with light scattering of scanning two aligned beads on slide. Those oscillation responses are related with deformability of chamber and the intrinsic characteristics of PZT. If we take nominal amplitude as actual amplitude for sinusoidal oscillations at 50 Hz, the amplitude is overestimated ~2 times in x-direction and ~3 times in y-direction. That will lead to huge errors for subsequent calibrations.

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2015 Tipo de documento: Article