Reduction of inappropriate anti-tachycardia pacing therapies and shocks by a novel suite of detection algorithms in heart failure patients with cardiac resynchronization therapy defibrillators: a historical comparison of a prospective database.
Europace
; 18(9): 1391-8, 2016 Sep.
Article
em En
| MEDLINE
| ID: mdl-26826135
AIMS: Implantable cardioverter defibrillators improve survival of patients at risk for ventricular arrhythmias, but inappropriate shocks occur in up to 30% of patients and have been associated with worse quality of life and prognosis. In heart failure patients with cardiac resynchronization therapy defibrillators (CRT-Ds), we evaluated whether a new generation of detection and discrimination algorithms reduces inappropriate shocks. METHODS AND RESULTS: We analysed 1983 Medtronic CRT-D patients (80% male, 67 ± 10 years), 1368 with standard devices (Control CRT-D) and 615 with new generation devices (New CRT-D). Expert electrophysiologists reviewed and classified the electrograms of all device-detected ventricular tachycardia/fibrillation episodes. Total follow-up was 3751 patients-years. Incidence of inappropriate shocks at 1 year was 2.8% [95% confidence interval (CI) = 2.0-3.5] in Control CRT-D and 0.9% (CI = 0.4-2.2) in New CRT-D (hazard ratio = 0.37, CI = 0.21-0.66, P < 0.001). In New CRT-D, inappropriate shocks were reduced by 77% [incidence rate ratio (IRR) = 0.23, CI = 0.16-0.35, P < 0.001] and inappropriate anti-tachycardia pacing by 81% (IRR = 0.19, CI = 0.11-0.335, P < 0.001). Annual rate per 100 patient-years for appropriate VF detections was 3.0 (CI = 2.1-4.2) in New CRT-D and 3.2 (CI = 2.1-5.0) in Control CRT-D (P = 0.68), for syncope was 0.4 (CI = 0.2-0.9) in New CRT-D and 0.7 (CI = 0.5-1.0) in Control CRT-D (P = 0.266), and for death was 1.0 (CI = 0.6-1.6) in New CRT-D and 3.5 (CI = 3.0-4.1) in Control CRT-D (P < 0.001). CONCLUSION: Detection and discrimination algorithms used in new generation CRT-D significantly reduced inappropriate shocks when compared with standard CRT-D. This result, with no compromise on VF sensitivity or risk of syncope, has important implications for patients' quality of life and prognosis.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Fibrilação Ventricular
/
Algoritmos
/
Processamento de Sinais Assistido por Computador
/
Falha de Prótese
/
Cardioversão Elétrica
/
Taquicardia Ventricular
/
Desfibriladores Implantáveis
/
Técnicas Eletrofisiológicas Cardíacas
/
Terapia de Ressincronização Cardíaca
/
Dispositivos de Terapia de Ressincronização Cardíaca
Tipo de estudo:
Clinical_trials
/
Diagnostic_studies
/
Etiology_studies
/
Observational_studies
/
Prognostic_studies
/
Risk_factors_studies
Limite:
Aged
/
Female
/
Humans
/
Male
/
Middle aged
País como assunto:
Europa
Idioma:
En
Ano de publicação:
2016
Tipo de documento:
Article