Your browser doesn't support javascript.
loading
Minimally invasive entero-enteral dual-path bypass using self-assembling magnets.
Ryou, Marvin; Aihara, Hiroyuki; Thompson, Christopher C.
Afiliação
  • Ryou M; Brigham and Women's Hospital, Harvard Medical School, Boston, MA, USA. mryou@partners.org.
  • Aihara H; Brigham and Women's Hospital, Harvard Medical School, Boston, MA, USA.
  • Thompson CC; Brigham and Women's Hospital, Harvard Medical School, Boston, MA, USA.
Surg Endosc ; 30(10): 4533-8, 2016 10.
Article em En | MEDLINE | ID: mdl-26895911
ABSTRACT

BACKGROUND:

A minimally invasive method of entero-enteral bypass may be desirable for treatment of obstruction, obesity, or metabolic syndrome. We have developed a technology based on miniature self-assembling magnets which create large-caliber anastomoses (incisionless anastomosis system or IAS). The aim of this study was to assess (a) procedural characteristics of IAS deployment and (b) long-term integrity and patency of the resulting jejuno-ileal dual-path bypass.

METHODS:

Endoscopic jejuno-ileal bypass creation using IAS magnets was performed in 8 Yorkshire pigs survived 3 months. PROCEDURE The jejunal magnet was endoscopically deployed. However, the ileal magnet required surgical delivery given restraints of porcine anatomy. A 5-mm enterotomy was created through which the ileal magnet was inserted using a modified laparoscopic delivery tool. Magnets were manually coupled. Pigs underwent serial endoscopies for anastomosis assessment. Three-month necropsies were performed, followed by pressure testing of anastomoses and histological analysis.

RESULTS:

Jejuno-ileal bypass creation using self-assembling IAS magnets was successful in all 8 pigs (100 %). Patent, leak-free bypasses formed in all animals by day 10. All IAS magnets were expelled by day 12. Anastomoses were widely patent at 3 months, with mean maximal diameter of 30 mm. At necropsy, adhesions were minimal. Pressure testing confirmed superior integrity of anastomotic tissue. Histology showed full epithelialization across the anastomosis with no evidence of submucosal fibrosis or inflammation.

CONCLUSIONS:

Entero-enteral bypass using self-assembling IAS magnets is safe and technically feasible in the porcine model. IAS magnets can be rapidly delivered endoscopically or through a modified laparoscopic device. Expulsion of fused magnets avoids retention of prosthetic material. Anastomoses are widely patent and fully re-epithelialized. Three-month pressure testing reveals anastomotic tissue to be as robust as native tissue, while necropsy and histology suggests minimal/absent tissue inflammation. In human anatomy, a fully endoscopic jejuno-ileal bypass using IAS magnets may be feasible.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Anastomose Cirúrgica / Derivação Jejunoileal / Desenho de Equipamento / Imãs / Íleo / Jejuno Tipo de estudo: Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Anastomose Cirúrgica / Derivação Jejunoileal / Desenho de Equipamento / Imãs / Íleo / Jejuno Tipo de estudo: Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 2016 Tipo de documento: Article