Your browser doesn't support javascript.
loading
Carbon-Ion Pencil Beam Scanning Treatment With Gated Markerless Tumor Tracking: An Analysis of Positional Accuracy.
Mori, Shinichiro; Karube, Masataka; Shirai, Toshiyuki; Tajiri, Minoru; Takekoshi, Takuro; Miki, Kentaro; Shiraishi, Yurika; Tanimoto, Katsuyuki; Shibayama, Kouichi; Yasuda, Shigeo; Yamamoto, Naoyoshi; Yamada, Shigeru; Tsuji, Hiroshi; Noda, Koji; Kamada, Tadashi.
Afiliação
  • Mori S; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan. Electronic address: shinshin@nirs.go.jp.
  • Karube M; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan.
  • Shirai T; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan.
  • Tajiri M; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan.
  • Takekoshi T; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan.
  • Miki K; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan.
  • Shiraishi Y; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan.
  • Tanimoto K; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan.
  • Shibayama K; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan.
  • Yasuda S; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan.
  • Yamamoto N; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan.
  • Yamada S; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan.
  • Tsuji H; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan.
  • Noda K; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan.
  • Kamada T; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan.
Int J Radiat Oncol Biol Phys ; 95(1): 258-266, 2016 May 01.
Article em En | MEDLINE | ID: mdl-26960747
ABSTRACT

PURPOSE:

Having implemented amplitude-based respiratory gating for scanned carbon-ion beam therapy, we sought to evaluate its effect on positional accuracy and throughput. METHODS AND MATERIALS A total of 10 patients with tumors of the lung and liver participated in the first clinical trials at our center. Treatment planning was conducted with 4-dimensional computed tomography (4DCT) under free-breathing conditions. The planning target volume (PTV) was calculated by adding a 2- to 3-mm setup margin outside the clinical target volume (CTV) within the gating window. The treatment beam was on when the CTV was within the PTV. Tumor position was detected in real time with a markerless tumor tracking system using paired x-ray fluoroscopic imaging units.

RESULTS:

The patient setup error (mean ± SD) was 1.1 ± 1.2 mm/0.6 ± 0.4°. The mean internal gating accuracy (95% confidence interval [CI]) was 0.5 mm. If external gating had been applied to this treatment, the mean gating accuracy (95% CI) would have been 4.1 mm. The fluoroscopic radiation doses (mean ± SD) were 23.7 ± 21.8 mGy per beam and less than 487.5 mGy total throughout the treatment course. The setup, preparation, and irradiation times (mean ± SD) were 8.9 ± 8.2 min, 9.5 ± 4.6 min, and 4.0 ± 2.4 min, respectively. The treatment room occupation time was 36.7 ± 67.5 min.

CONCLUSIONS:

Internal gating had a much higher accuracy than external gating. By the addition of a setup margin of 2 to 3 mm, internal gating positional error was less than 2.2 mm at 95% CI.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Radioterapia com Íons Pesados / Neoplasias Hepáticas / Neoplasias Pulmonares / Movimento Tipo de estudo: Evaluation_studies / Observational_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Radioterapia com Íons Pesados / Neoplasias Hepáticas / Neoplasias Pulmonares / Movimento Tipo de estudo: Evaluation_studies / Observational_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2016 Tipo de documento: Article