Evaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film.
Sonier, Marcus; Wronski, Matt; Yeboah, Collins.
Afiliação
  • Sonier M; Sunnybrook Health Sciences Centre Odette Cancer Centre. marcus.sonier@sunnybrook.ca.
J Appl Clin Med Phys ; 17(2): 304-314, 2015 03 08.
Article em En | MEDLINE | ID: mdl-27074448
ABSTRACT
Lens dose is a concern during the treatment of facial lesions with anterior electron beams. Lead shielding is routinely employed to reduce lens dose and minimize late complications. The purpose of this work is twofold 1) to measure dose pro-files under large-area lead shielding at the lens depth for clinical electron energies via film dosimetry; and 2) to assess the accuracy of the Pinnacle treatment planning system in calculating doses under lead shields. First, to simulate the clinical geometry, EBT3 film and 4 cm wide lead shields were incorporated into a Solid Water phantom. With the lead shield inside the phantom, the film was positioned at a depth of 0.7 cm below the lead, while a variable thickness of solid water, simulating bolus, was placed on top. This geometry was reproduced in Pinnacle to calculate dose profiles using the pencil beam electron algorithm. The measured and calculated dose profiles were normalized to the central-axis dose maximum in a homogeneous phantom with no lead shielding. The resulting measured profiles, functions of bolus thickness and incident electron energy, can be used to estimate the lens dose under various clinical scenarios. These profiles showed a minimum lead margin of 0.5 cm beyond the lens boundary is required to shield the lens to ≤ 10% of the dose maximum. Comparisons with Pinnacle showed a consistent overestimation of dose under the lead shield with discrepancies of ~ 25% occur-ring near the shield edge. This discrepancy was found to increase with electron energy and bolus thickness and decrease with distance from the lead edge. Thus, the Pinnacle electron algorithm is not recommended for estimating lens dose in this situation. The film measurements, however, allow for a reasonable estimate of lens dose from electron beams and for clinicians to assess the lead margin required to reduce the lens dose to an acceptable level.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Algoritmos / Imagens de Fantasmas / Elétrons / Dosimetria Fotográfica / Cristalino Tipo de estudo: Evaluation_studies / Health_economic_evaluation Limite: Humans Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Algoritmos / Imagens de Fantasmas / Elétrons / Dosimetria Fotográfica / Cristalino Tipo de estudo: Evaluation_studies / Health_economic_evaluation Limite: Humans Idioma: En Ano de publicação: 2015 Tipo de documento: Article