A snare in time--adding a safety mechanism to the Amplatzer Septal Occluder.
Latson, Larry A.
Afiliação
  • Latson LA; Clinical Professor of Pediatrics, Florida Atlantic University.
Catheter Cardiovasc Interv ; 87(5): 931-2, 2016 Apr.
Article em En | MEDLINE | ID: mdl-27085031
Delivery cable/attachment of the Amplatzer Septal Occluder holds the device at a suboptimal angle relative to the plane of the atrial septum prior to release A small gooseneck snare tightened on the hub can maintain flexible control of the device after release of the delivery cable This technique can improve safety of delivery in difficult cases if you think ahead.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Dispositivo para Oclusão Septal / Comunicação Interatrial Limite: Humans Idioma: En Ano de publicação: 2016 Tipo de documento: Article