Your browser doesn't support javascript.
loading
Aspirex Thrombectomy in Occluded Dialysis Access: A Retrospective Study.
Dyer, Jules; Rosa, Joao; Chachlani, Menka; Nicholas, Johann.
Afiliação
  • Dyer J; Department of Radiology, New Cross Hospital, Wolverhampton, West Midlands, WV10 0QP, UK. Jules.Dyer@nhs.net.
  • Rosa J; ST1 Radiology, University Hospitals Bristol NHS Foundation Trust, Bristol, UK.
  • Chachlani M; University of Birmingham, Birmingham, West Midlands, UK.
  • Nicholas J; Department of Radiology, New Cross Hospital, Wolverhampton, West Midlands, WV10 0QP, UK.
Cardiovasc Intervent Radiol ; 39(10): 1484-90, 2016 Oct.
Article em En | MEDLINE | ID: mdl-27094693
ABSTRACT

PURPOSE:

This study is the first to present the outcomes of the Straub Aspirex device for the salvage of occluded renal dialysis access fistulae. MATERIALS AND

METHODS:

This is a retrospective study, using data from the Renal Unit and Radiology Department database. It included all the patients between 2010 and 2014 who underwent percutaneous mechanical thrombectomy (PMT) treated by JD. Aspirex is an over-the-wire, 6-10 French catheter within which is a rapidly rotating helix which draws thrombus into a window near the tip which it then macerates and removes. Access survival was assessed using the Kaplan-Meier method, and multi-variant analysis was performed using the Cox proportional hazards model. Significance was considered if p < 0.05.

RESULTS:

A total of 27 procedures were performed for 19 patients. 13 had autologous arterio-venous fistulae, and 14 had synthetic (PTFE) arterio-venous grafts. 15 were males, 4 females. 100 % of the patients successfully had a channel of thrombus removed. This resulted in an 81.5 % initial clinical success, with primary patency rates of 53.6, 44.3 and 33 % by days 30, 90 and 480, respectively, without significant difference of any analysed covariates. No major complication (pulmonary embolus, paradoxical cerebral infarction, limb ischaemia or significant haemorrhage) occurred.

CONCLUSION:

Aspirex has rates of patency and complication similar to other PMT devices. No covariant studied affected outcome.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Sucção / Trombose / Derivação Arteriovenosa Cirúrgica / Diálise Renal / Trombectomia / Desenho de Equipamento / Oclusão de Enxerto Vascular Tipo de estudo: Observational_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Sucção / Trombose / Derivação Arteriovenosa Cirúrgica / Diálise Renal / Trombectomia / Desenho de Equipamento / Oclusão de Enxerto Vascular Tipo de estudo: Observational_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2016 Tipo de documento: Article