Your browser doesn't support javascript.
loading
A Novel Free Flap Monitoring System Using Tissue Oximetry with Text Message Alerts.
Ricci, Joseph A; Vargas, Christina R; Lin, Samuel J; Tobias, Adam M; Taghinia, Amir H; Lee, Bernard T.
Afiliação
  • Ricci JA; Division of Plastic and Reconstructive Surgery, Beth Israel Deaconess Medical Center, Boston, Massachusetts.
  • Vargas CR; Division of Plastic and Reconstructive Surgery, Beth Israel Deaconess Medical Center, Boston, Massachusetts.
  • Lin SJ; Division of Plastic and Reconstructive Surgery, Beth Israel Deaconess Medical Center, Boston, Massachusetts.
  • Tobias AM; Division of Plastic and Reconstructive Surgery, Beth Israel Deaconess Medical Center, Boston, Massachusetts.
  • Taghinia AH; Department of Plastic Surgery, Boston Children's Hospital, Boston, Massachusetts.
  • Lee BT; Division of Plastic and Reconstructive Surgery, Beth Israel Deaconess Medical Center, Boston, Massachusetts.
J Reconstr Microsurg ; 32(5): 415-20, 2016 Jun.
Article em En | MEDLINE | ID: mdl-27135144
ABSTRACT
Background The time to detection of vascular compromise is a significant predictor of free flap salvage outcomes as early reexploration improves salvage rates. Continuous transcutaneous near-infrared tissue oximetry is an objective, quantitative method of detecting flap vascular compromise and has been shown to allow earlier reexploration and higher salvage rates than clinical assessment alone. We designed a novel text messaging system to improve communication using tissue oximetry monitoring. Methods A retrospective review was performed of a prospectively collected database of all microsurgical breast reconstructions from 2008 to 2015. A novel text messaging system was introduced in 2013 and programmed to send text messages alert when the tissue oximetry readings suggested potential flap compromise based on established thresholds. Patient demographics and complications, including rate of reexploration and flap loss were assessed. Results There were 900 autologous microsurgical breast free flaps during the study period 614 were monitored with standard clinical monitoring and tissue oximetry compared with 286 flaps with the additional text messaging system. There were 27 unplanned returns to the operating room in the tissue oximetry group and 5 in the text messaging group with 1 complete flap loss in each group. Reexploration occurred sooner as a result of these text message alerts (17.5 vs. 26.6 hours postoperatively), however, it did not achieve statistical significance. Conclusions We were able to demonstrate the use of a novel text messaging system for tissue oximetry. This alert system shows promise in identifying impending flap loss with rapid notification of the surgical team.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Reoperação / Oximetria / Mamoplastia / Envio de Mensagens de Texto / Microcirurgia / Monitorização Fisiológica Tipo de estudo: Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Female / Humans / Middle aged Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Reoperação / Oximetria / Mamoplastia / Envio de Mensagens de Texto / Microcirurgia / Monitorização Fisiológica Tipo de estudo: Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Female / Humans / Middle aged Idioma: En Ano de publicação: 2016 Tipo de documento: Article