Your browser doesn't support javascript.
loading
Spontaneous Sealing of a Type Ia Endoleak after Ovation Stent Graft Implantation in a Patient with On-Label Aortic Neck Anatomy.
Pranteda, Chiara; Sirignano, Pasqualino; Capoccia, Laura; Mansour, Wassim; Montelione, Nunzio; Speziale, Francesco.
Afiliação
  • Pranteda C; Vascular and Endovascular Surgery Division, Department of Surgery "Paride Stefanini", Policlinico Umberto I, "Sapienza" University of Rome, Rome, Italy. Electronic address: chiara.pranteda@gmail.com.
  • Sirignano P; Vascular and Endovascular Surgery Division, Department of Surgery "Paride Stefanini", Policlinico Umberto I, "Sapienza" University of Rome, Rome, Italy.
  • Capoccia L; Vascular and Endovascular Surgery Division, Department of Surgery "Paride Stefanini", Policlinico Umberto I, "Sapienza" University of Rome, Rome, Italy.
  • Mansour W; Vascular and Endovascular Surgery Division, Department of Surgery "Paride Stefanini", Policlinico Umberto I, "Sapienza" University of Rome, Rome, Italy.
  • Montelione N; Vascular and Endovascular Surgery Division, Department of Surgery "Paride Stefanini", Policlinico Umberto I, "Sapienza" University of Rome, Rome, Italy.
  • Speziale F; Vascular and Endovascular Surgery Division, Department of Surgery "Paride Stefanini", Policlinico Umberto I, "Sapienza" University of Rome, Rome, Italy.
Ann Vasc Surg ; 34: 270.e19-24, 2016 Jul.
Article em En | MEDLINE | ID: mdl-27174349
ABSTRACT
We report a case of an early type Ia endoleak after endovascular aneurysm repair (EVAR) of an abdominal aortic aneurysm by Ovation Stent Graft implantation and spontaneously resolved without further reintervention. The patient presents a conical aortic neck, but EVAR was performed within the instruction for use proposed by manufactory. At completion angiography, a low-flow type Ia endoleak was present and left untreated. Computed tomographic angiography performed on the third postoperative day showed infolding of the 2 sealing rings. The patient was dismissed without further treatment. At 3-month follow-up, the leak appeared spontaneously sealed with partial expansion of the 2 rings.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Prótese Vascular / Stents / Aneurisma da Aorta Abdominal / Implante de Prótese Vascular / Endoleak / Procedimentos Endovasculares Limite: Aged / Humans / Male Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Prótese Vascular / Stents / Aneurisma da Aorta Abdominal / Implante de Prótese Vascular / Endoleak / Procedimentos Endovasculares Limite: Aged / Humans / Male Idioma: En Ano de publicação: 2016 Tipo de documento: Article