Your browser doesn't support javascript.
loading
Observations on failed retrieval of optional inferior vena cava filters.
Ballard, David H; Do, Daniel V; Laborde, Jeremy J; de Gregorio, Miguel A; Shi, Runhua; D'Agostino, Horacio B.
Afiliação
  • Ballard DH; Department of Surgery, Louisiana State University Health Shreveport, 1501 Kings Highway, Shreveport, LA 71103, USA. Electronic address: davidballardmd@gmail.com.
  • Do DV; Department of Radiology, Louisiana State University Health Shreveport, 1501 Kings Highway, Shreveport, LA 71103, USA.
  • Laborde JJ; Department of Radiology, Louisiana State University Health Shreveport, 1501 Kings Highway, Shreveport, LA 71103, USA.
  • de Gregorio MA; Department of Interventional Radiology, University of Zaragoza, Gomez Laguna, 13, 5 B, 50009 Saragossa, Spain.
  • Shi R; Department of Medicine and Feist-Weiller Cancer Center, LSU Health Shreveport, 1501 Kings Hwy, Shreveport, LA 71103, USA.
  • D'Agostino HB; Department of Radiology, Louisiana State University Health Shreveport, 1501 Kings Highway, Shreveport, LA 71103, USA. Electronic address: hdagos@lsuhsc.edu.
Clin Imaging ; 40(5): 931-5, 2016.
Article em En | MEDLINE | ID: mdl-27196734
ABSTRACT

PURPOSE:

To evaluate causes of failed optional inferior vena cava filter (IVCF) retrievals.

METHODS:

Single-center retrospective study.

RESULTS:

IVCF retrievals were attempted in 26/211 (12%) patients at a mean 42.9 days. There were 9 failures (all OptEase) due to inability to snare the hook (n=5), noncollapsible IVCF (n=3), and unusual procedural pain (n=1). Median duration of retrieved IVCFs was 31 days compared to 53 days for failures (P<.05). IVCFs aligned with the IVC's cephalocaudal axis were retrieved in 13/16 cases, while misaligned IVCFs were retrieved 4/10 cases (P<.05).

CONCLUSION:

Filter duration and misalignment were significantly associated with retrieval failures.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Embolia Pulmonar / Filtros de Veia Cava / Procedimentos Endovasculares Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Female / Humans / Male / Middle aged País como assunto: America do norte Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Embolia Pulmonar / Filtros de Veia Cava / Procedimentos Endovasculares Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Female / Humans / Male / Middle aged País como assunto: America do norte Idioma: En Ano de publicação: 2016 Tipo de documento: Article