A Multiplex Electrochemical Biosensor for Bloodstream Infection Diagnosis.
Gao, Jian; Jeffries, Lindsie; Mach, Kathleen E; Craft, David W; Thomas, Neal J; Gau, Vincent; Liao, Joseph C; Wong, Pak Kin.
Afiliação
  • Gao J; 1 Department of Biomedical Engineering, Pennsylvania State University, University Park, PA, USA.
  • Jeffries L; 2 Department of Biomedical Engineering, The University of Arizona, Tucson, AZ, USA.
  • Mach KE; 3 Department of Urology, Stanford University School of Medicine, Stanford, CA, USA.
  • Craft DW; 4 Pathology and Laboratory Medicine, Penn State Milton S. Hershey Medical Center, Hershey, PA, USA.
  • Thomas NJ; 5 Departments of Pediatrics and Public Health Sciences, Penn State University College of Medicine, Hershey, PA, USA.
  • Gau V; 6 GeneFluidics, Inc., Irwindale, CA, USA.
  • Liao JC; 3 Department of Urology, Stanford University School of Medicine, Stanford, CA, USA.
  • Wong PK; 1 Department of Biomedical Engineering, Pennsylvania State University, University Park, PA, USA.
SLAS Technol ; 22(4): 466-474, 2017 08.
Article em En | MEDLINE | ID: mdl-27226118
Accurate and timely detection of bacterial pathogens will improve the clinical management of infections. Herein, we demonstrate an electrochemical biosensor that directly detects bacteria in human blood samples, resulting in the rapid diagnosis of a bloodstream infection. The multiplex biosensor detects the species-specific sequences of the 16S ribosomal RNA of bacteria for pathogen identification in physiological samples without preamplification. The analytical performance characteristics of the biosensor, including the limit of detection and probe cross-reactivity, are evaluated systematically. The feasibility of the biosensor for a diagnosis of a bloodstream infection is demonstrated by identifying bacterial clinical isolates spiked in whole blood and blood culture samples that were tested positive for bacteria. The electrochemical biosensor correctly identifies all the species in the samples with 100% concordance to microbiological analysis.