Your browser doesn't support javascript.
loading
Comparison of pre-processing methods for multiplex bead-based immunoassays.
Rausch, Tanja K; Schillert, Arne; Ziegler, Andreas; Lüking, Angelika; Zucht, Hans-Dieter; Schulz-Knappe, Peter.
Afiliação
  • Rausch TK; Institut für Medizinische Biometrie und Statistik, Universität zu Lübeck, Universitätsklinikum Schleswig-Holstein, Campus Lübeck, Ratzeburger Allee 160, 23562, Lübeck, Germany.
  • Schillert A; Institut für Medizinische Biometrie und Statistik, Universität zu Lübeck, Universitätsklinikum Schleswig-Holstein, Campus Lübeck, Ratzeburger Allee 160, 23562, Lübeck, Germany.
  • Ziegler A; Institut für Medizinische Biometrie und Statistik, Universität zu Lübeck, Universitätsklinikum Schleswig-Holstein, Campus Lübeck, Ratzeburger Allee 160, 23562, Lübeck, Germany. ziegler@imbs.uni-luebeck.de.
  • Lüking A; Zentrum für Klinische Studien, Universität zu Lübeck, Lübeck, Germany. ziegler@imbs.uni-luebeck.de.
  • Zucht HD; School of Mathematics, Statistics and Computer Science, University of KwaZulu-Natal, Pietermaritzburg, South Africa. ziegler@imbs.uni-luebeck.de.
  • Schulz-Knappe P; Protagen AG, Dortmund, Germany.
BMC Genomics ; 17(1): 601, 2016 Aug 11.
Article em En | MEDLINE | ID: mdl-27515389
ABSTRACT

BACKGROUND:

High throughput protein expression studies can be performed using bead-based protein immunoassays, such as the Luminex® xMAP® technology. Technical variability is inherent to these experiments and may lead to systematic bias and reduced power. To reduce technical variability, data pre-processing is performed. However, no recommendations exist for the pre-processing of Luminex® xMAP® data.

RESULTS:

We compared 37 different data pre-processing combinations of transformation and normalization methods in 42 samples on 384 analytes obtained from a multiplex immunoassay based on the Luminex® xMAP® technology. We evaluated the performance of each pre-processing approach with 6 different performance criteria. Three performance criteria were plots. All plots were evaluated by 15 independent and blinded readers. Four different combinations of transformation and normalization methods performed well as pre-processing procedure for this bead-based protein immunoassay.

CONCLUSIONS:

The following combinations of transformation and normalization were suitable for pre-processing Luminex® xMAP® data in this study weighted Box-Cox followed by quantile or robust spline normalization (rsn), asinh transformation followed by loess normalization and Box-Cox followed by rsn.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Autoanticorpos / Imunoensaio / Regulação da Expressão Gênica / Perfilação da Expressão Gênica Limite: Humans Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Autoanticorpos / Imunoensaio / Regulação da Expressão Gênica / Perfilação da Expressão Gênica Limite: Humans Idioma: En Ano de publicação: 2016 Tipo de documento: Article