Your browser doesn't support javascript.
loading
Optimization of Cellular analysis of Synovial Fluids by optical microscopy and automated count using the Sysmex XN Body Fluid Mode.
Seghezzi, Michela; Buoro, Sabrina; Manenti, Barbara; Mecca, Tommaso; Ferrari, Roberto; Zappalà, Giorgio; Castelli, Claudio Carlo; Balboni, Fiamma; Pezzati, Paola; Ottomano, Cosimo; Lippi, Giuseppe.
Afiliação
  • Seghezzi M; Clinical Chemistry Laboratory, Papa Giovanni XXIII Hospital, Bergamo, Italy.
  • Buoro S; Clinical Chemistry Laboratory, Papa Giovanni XXIII Hospital, Bergamo, Italy. Electronic address: sbuoro@asst-pg23.it.
  • Manenti B; Clinical Chemistry Laboratory, Papa Giovanni XXIII Hospital, Bergamo, Italy.
  • Mecca T; Clinical Chemistry Laboratory, Papa Giovanni XXIII Hospital, Bergamo, Italy.
  • Ferrari R; Orthopedics and Traumatology Unit. Hospital Papa Giovanni XXIII, Bergamo, Italy.
  • Zappalà G; Orthopedics and Traumatology Unit. Hospital Papa Giovanni XXIII, Bergamo, Italy.
  • Castelli CC; Orthopedics and Traumatology Unit. Hospital Papa Giovanni XXIII, Bergamo, Italy.
  • Balboni F; Clinical Chemistry Laboratory, Istituto Fiorentino di Cura e Assistenza (IFCA), Firenze, Italy.
  • Pezzati P; Clinical Chemistry Laboratory Hospital Careggi Firenze, , University of Firenze, Italy.
  • Ottomano C; Synlab, Monza, Italy.
  • Lippi G; Section of Clinical Biochemistry, University of Verona, Verona, Italy.
Clin Chim Acta ; 462: 41-48, 2016 Nov 01.
Article em En | MEDLINE | ID: mdl-27581597
ABSTRACT

BACKGROUND:

This study was planned to assess the impact of pre-treating synovial fluid (SF) samples with hyaluronidase (HY), defining the best procedure for optical microscopy (OM) analysis and evaluating the performance of Sysmex XN-9000 Body Fluid module (XN-BF).

METHODS:

The cell count by OM was carried out both with and without HY pre-treatment, and using 3 different types of staining reagents. The evaluation of XN-BF included data comparison with OM (100 SFs), carryover, Limit of Blank (LoB), Limit of Detection (LoD), Limit of Quantitation (LoQ) and linearity.

RESULTS:

Unlike cell count in Burker's chamber and staining with Stromatol, pre-treatment with HY and staining with Methylene Blue and Turk's promoted cell clustering. The SF samples pre-treated with HY displayed excellent morphological quality, contrary to samples without HY pre-treatment. Excellent correlation was found between total cells counting with both OM and XN-BF. Satisfactory agreement was also observed between polymorphonuclear neutrophils compared to XN-BF parameter, whereas mononuclear cell count on XN-BF had suboptimal agreement with OM. The carryover was negligible. The LoB, LoD, LoQ and linearity were excellent.

CONCLUSION:

XN-BF displays excellent performance, which makes it a reliable and practical alternative to OM for SF samples analysis in clinical laboratories.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Líquido Sinovial / Contagem de Células / Automação Laboratorial / Citometria de Fluxo Limite: Humans Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Líquido Sinovial / Contagem de Células / Automação Laboratorial / Citometria de Fluxo Limite: Humans Idioma: En Ano de publicação: 2016 Tipo de documento: Article