Comparison of the processed electroencephalogram and awake neurological assessment during carotid endarterectomy.
Anaesth Intensive Care
; 17(3): 298-304, 1989 Aug.
Article
em En
| MEDLINE
| ID: mdl-2774148
A processed electroencephalogram (EEG) produced by the Lifescan monitor (Neurometrics), was compared to awake neurological assessment for detecting cerebral ischaemia in seventy patients undergoing carotid endarterectomy under cervical plexus block. Of the six patients demonstrating neurological signs on cross-clamping the carotid, five displayed simultaneous EEG changes, four being detected during surgery, and one being detected after reviewing the EEG postoperatively. Another four patients displayed EEG changes indicative of ischaemia but unassociated with neurological signs. A further patient displayed contralateral intraoperative EEG changes. Hypotension resulted in one EEG change and two cases were associated with technical difficulties with the monitor. The presence of false negatives, possible false positives, technical errors and subjective interpretation associated with the processed EEG make it less reliable than awake neurological assessment for the detection of cerebral ischaemia.