Voltage-controlled accommodating IOL system using an ion polymer metal composite actuator.
Opt Express ; 24(20): 23280-23288, 2016 Oct 03.
Article em En | MEDLINE | ID: mdl-27828392
ABSTRACT
Surgeons treat cataracts by replacing the clouded lens with an intraocular lens (IOL), but patients are required to wear reading glasses for tasks requiring near vision. We suggest a new voltage-controlled accommodating IOL made of an ionic polymer metal composite (IPMC) actuator to change focus. An in vitro experiment was conducted where an actuator was placed inside the eye and moved with applied voltage. The lens attached to the actuator was deformed by its movement to change the patient's focus. The results showed that this system can accommodate a change of approximately 0.8 diopters under an applied voltage of ± 1.3 V.